Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1809-2024Class II

LASER Application Instrument, 23 cm, REF: 8694; LEUNIG-GREVERS LASER Application, REF: 461000

Karl Storz Endoscopy-America, Inc. / initiated 2024-04-01 / Open, Classified / root cause: design

Karl Storz Endoscopy began a recall of LASER Application Instrument, 23 cm, REF: 8694; LEUNIG-GREVERS LASER Application, REF: 461000 on . FDA classified it as Class II and gives the reason as "Laser surgical instruments have no specific evidence of a validated reprocessing method, therefore these instruments lack sterility assurance, so once the instruments have gone through reprocessing after use and then...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1809-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Karl Storz Endoscopy-America, Inc.
Product
LASER Application Instrument, 23 cm, REF: 8694; LEUNIG-GREVERS LASER Application, REF: 461000
Product code
GEX Powered Laser Surgical Instrument
Reason for recall
Laser surgical instruments have no specific evidence of a validated reprocessing method, therefore these instruments lack sterility assurance, so once the instruments have gone through reprocessing after use and then are used again, there is an increased risk of the patient being exposed to infection.
Root cause tag
design
FDA root cause
Device Design
Action
On 4/1/24, recall notices were mailed to representatives for medical product safety, users, operators, distributors who were asked to do the following: 1) Quarantine and discontinue use of affected devices. 2) Pass on the recall notice to all users of the affected devices and to all other persons who need to be aware within your organization. 3) If you have or may have distributed the products listed, please identify and promptly notify those recipients, or provide the recalling firm a list of customers who received/may have received the products listed. 4) Complete and return the customer reply form via email to [email protected] 5) If you have product to return contact the firm at the following phone numbers: For Human Product Telephone: 800-421-0837 Option 1 For Vet Product Telephone: 800-955-7832 Option 1 The firm has placed all affected product on hold and has ceased the distribution of the affected product in the U.S. until further notice. For all other questions regarding this communication, please send an email to [email protected]
Quantity in commerce
85
Distribution
Worldwide distribution - US Nationwide in the states of OH, CA, FL, MA, TX, KY, WI, MI, MN, AR, NE, IL, NY, HI and the country of Argentina.
Product codes and lots
All lots. REF/UDI-DI: 8694/04048551216392, 461000/04048551164860

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1809-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1809-2024Class IIOngoingVoluntary: Firm initiated