Z-1831-2024Class II

CARDIOHELP-i System. Blood oxygenation and carbon dioxide removal system

Maquet Cardiovascular Us Sales, LLC / initiated 2024-04-15 / Open, Classified

Maquet Medical Systems USA began a recall of CARDIOHELP-i System. Blood oxygenation and carbon dioxide removal system on . FDA classified it as Class II and gives the reason as "The firm identified that the measured patient leakage current in some devices significantly exceeded expected levels". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1831-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
CARDIOHELP-i System. Blood oxygenation and carbon dioxide removal system.
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
The firm identified that the measured patient leakage current in some devices significantly exceeded expected levels.
FDA root cause
Under Investigation by firm
Action
Consignees received an "URGENT MEDICAL DEVICE - Removal" email, dated 4/15/2024, notifying them of this recall event. Consignees are asked to examine their inventory to identify if they possess any affected devices and all consignees are asked to return the provided Response Form via email to identify where affected devices are located. Consignees that report having any affected devices will be contacted by Getinge to arrange for return of the affected device to upgrade the system's sensor panel and perform testing to ensure that leakage current is within specifications. Technical Support can also be contacted at 888-943-8872 (option 4, option 2, option 1) Monday through Friday between 8:00 AM and 6:00 PM EST to make service arrangements. Consignees are asked to not use affected devices for any additional patients until it has been cleared by the MCP/Getinge system check, repaired, or replaced. The provided recall notice is to be forwarded to users of the CARDIOHELP-i system and if the device was further distributed.
Quantity in commerce
148 units (US: 5; OUS: 143)
Distribution
US Nationwide distribution in the states of NJ & TX.
Product codes and lots
Part No. 70104.8012 and 70104.8012R; UDI-DI: 04037691658384; Serial no. 90410021, 90410029, 90410165, 90410254, 90410255, 90410024, 90410025, 90410026, 90410028, 90410030, 90410031, 90410034, 90410035, 90410036, 90410043, 90410044, 90410045, 90410046, 90410047, 90410048, 90410049, 90410050, 90410051, 90410052, 90410053, 90410054, 90410055, 90410056, 90410057, 90410058, 90410059, 90410060, 90410061, 90410062, 90410063, 90410064, 90410065, 90410067, 90410068, 90410069, 90410070, 90410071, 90410072, 90410073, 90410074, 90410075, 90410076, 90410077, 90410078, 90410079, 90410080, 90410081, 90410082, 90410084, 90410085, 90410086, 90410087, 90410088, 90410089, 90410091, 90410093, 90410094, 90410095, 90410096, 90410097, 90410098, 90410099, 90410100, 90410101, 90410102, 90410103, 90410104, 90410105, 90410106, 90410107, 90410108, 90410109, 90410110, 90410111, 90410112, 90410113, 90410114, 90410115, 90410116, 90410117, 90410118, 90410119, 90410120, 90410121, 90410122, 90410123, 90410124, 90410125, 90410126, 90410127, 90410130, 90410131, 90410132, 90410133, 90410134, 90410135, 90410136, 90410137, 90410138, 90410139, 90410140, 90410141, 90410142, 90410143, 90410144, 90410146, 90410147, 90410148, 90410149, 90410151, 90410152, 90410153, 90410154, 90410155, 90410156, 90410157, 90410158, 90410159, 90410160, 90410161, 90410162, 90410163, 90410166, 90410167, 90410168, 90410169, 90410170, 90410172, 90410173, 90410174, 90410175, 90410176, 90410177, 90410178, 90410179, 90410180, 90410181, 90410182, 90410183, 90410184, 90410185, 90410186, 90410187.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1831-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1831-2024Class IIOngoingVoluntary: Firm initiated