Ingenia 1.5T with Evolution Upgrade 1.5T - Magnetic Resonance (MR) systems are Medical Electrical Systems...
Philips North America Llc began a recall of Ingenia 1.5T with Evolution Upgrade 1.5T - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device REF: 782116 on . FDA classified it as Class II and gives the reason as "Philips Ingenia Evolution systems with software R5.7.1 may not provide warning for low ventilation setting which may lead to increase in body temperature or burning sensation". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1829-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips North America LLC
- Product
- Ingenia 1.5T with Evolution Upgrade 1.5T - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device REF: 782116
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- Philips Ingenia Evolution systems with software R5.7.1 may not provide warning for low ventilation setting which may lead to increase in body temperature or burning sensation
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Philips issued URGENT Medical Device Correction letter on 4/22/24 via FedEx. Philips Markets Organizations are responsible for distributing the letters outside of the U.S. Letter states reason for recall, health risk and action to take: a. Customers can continue to use the identified system(s) in accordance with the intended use pertaining to setting patient ventilation. (Reference IFU, including below information). {If your Ingenia 1.5T or Ingenia 3.0T system(s) has been upgraded to Ingenia Evolution configuration and running R5.7 software, it is impacted by this issue. (Note: customer will not see SW version detail of R5.7.1). o As stated in both IFU sections Adjust Ventilation in Bore and Adjusting the Bore Ventilation, The MR system has a recommended ventilation level. This is a static level based on average scans and system specific conditions. The system does not change the recommended ventilation level for e.g. high SAR scans, examination room temperature and patient weights, clothing and conditions. It is the responsibility of the operator to determine the level of ventilation for different scans and conditions together with the patient. Note: The recommended level is identified as level 3 in the IFU and in the images shown in Figure 1. o When setting ventilation at the VitalScreen, (IFU section Adjusting the Bore Ventilation) and ventilation is set to a value marked with a caution sign, you are warned that the patient might get too warm and that you need to pay attention. o When setting ventilation at the Operator Console (IFU Section Adjusting Ventilation in Bore): Current patient ventilation level, Level 3 is recommended. Use the <=> and <-> buttons above to modify the patient ventilation level. o Note: Should Patient Ventilation level be below level 3, operator is to pay additional attention to patient temperature. b. Refer to Appendix A for specific details regarding the software issue descriptions and advice to customers. c. Please co
- Quantity in commerce
- 32 units
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, KS, MS, NH and the countries of Austria, Croatia, Finland, France, Germany, Italy, Lebanon, Netherlands, Norway, Sweden, Switzerland, Thailand, United Kingdom.
- Product codes and lots
- UDI-DI: (01)00884838099722(21) Serial Numbers: 41017 41024 41025 41041 41047 41051 41080 41089 41130 41143 41144 41189 41196 41226 41235 41286 41287 41344 41367 41413 41432 41433 41497 41520 70093 70099 70202 70260 70589 70996 81073 84370
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1829-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1830-2024 | Class II | Ongoing | Voluntary: Firm initiated |