Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2130-2024Class I

ZOLL 731 Ventilator (EMV+, AEV, Eagle II) - for MRI Compatible Devices - Operator's Guide and Quick Reference...

ZOLL Medical Corporation / initiated 2024-04-30 / Open, Classified

ZOLL Medical Corporation began a recall of ZOLL 731 Ventilator (EMV+, AEV, Eagle II) - for MRI Compatible Devices - Operator's Guide and Quick Reference Guide (QRGs) as follows: Part Numbers: 1... on . FDA classified it as Class I and gives the reason as "Operator's Guide & Quick Reference Guide (QRG) ZOLL 731 Ventilator for MRI Compatible Devices. Patient Safety Ventilator MRI information omitted from the manuals. Update to the instructions ensures safe distance from...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2130-2024
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
ZOLL Medical Corporation
Product
ZOLL 731 Ventilator (EMV+, AEV, Eagle II) - for MRI Compatible Devices - Operator's Guide and Quick Reference Guide (QRGs) as follows: Part Numbers: 1. 799-EGL2-02 2. 799-EGL2-02-01 3. 799-EGL2-02-04 4. 799-EGL2-02-05 5. 799-EGL2-02-08 6. 799-EGL2-02-10 7. 799-EGL2-02-11 8. 799-EGL2-02-15 9. 799-EGL2-02-16 10. 799-EMBP-05-01 11. 799-EMBP-05-02 12. 799-EMBP-05-08 13. 799-EMBP-05-10 14. 799-EMBP-05-11 15. 799-EMBP-05-12 16. 799-EMBP-05-12-68 17. 799-EMBP-05-16 18. 799-EMBP-05-23 19. 799-EMVP-05 20. 799-EMVP-05-01 21. 799-EMVP-05-05 22. 799-EMVP-05-10 23. 799-EMVP-05-15 24. 799-EMVP-05-16 Manual Part Numbers 906-0731-01; 9650-002365-01 (US); 906-0731-01-05 (CAN); 906-0731-01-10; 9650-0002363-01; 906-0731-01 (Commercial); 906-0731-03 (Military); 9650-0002363-01; 9650-0002363-05; 906-0731-01; 906-0731-01-05 (CAN); 906-0731-01-04; 906-0731-01-05; 906-0731-01-08; 906-0731-01-10; 906-0731-01-11; 906-0731-01-15; 906-0731-01-16; 9650-0002363-01; 9650-0002363-02; 9650-0002363-08; 9650-0002363-10; 9650-0002363-11; 9650-0002363-12; 9650-0002363-16; 9650-0002363-23; 906-0731-01 (Commercial); 906-0731-03 (Military); 9650-0002363-01; 9650-0002363-05; 9650-0002363-10; 9650-0002363-15; 9650-0002363-16 Quick Reference Guide (QRG) Part Numbers: 907-0731-04; 9652-000499-01; 9652-000511-01; 907-0731-04; 9652-000511-01; 907-0731-04; 9652-000499-01; 9652-000499-04; 9652-000499-08; 9652-000499-10; 9652-000499-11; 907-0731-04; 9652-000499-16; 9652-000511-01; 9652-000511-02; 9652-000511-08; 9652-000511-10; 9652-000511-11; 9652-000511-12; 9652-000511-16; 9652-000511-23; 907-0731-04; 9652-000511-01; 9652-000511-05; 9652-000511-10; 9652-000511-15; 9652-000511-16
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
Operator's Guide & Quick Reference Guide (QRG) ZOLL 731 Ventilator for MRI Compatible Devices. Patient Safety Ventilator MRI information omitted from the manuals. Update to the instructions ensures safe distance from the MRI to ensure proper function of the ventilator and reduce the risk of delay in ventilator therapy
FDA root cause
Under Investigation by firm
Action
Zoll issued Urgent Corrective Action Letter dated 4/22/24. Letter states reason for recall, health, risk and action to take: Customers who have affected manuals should take the following steps: (1) Locate all old versions of affected Operator s Guides and/or Quick Reference Guides. (2) Complete and return the attached customer response form via e-mail, fax, or mail to request replacement manuals. (3) Alert all users of the ZOLL Ventilator - with MRI Compatibility, to ensure they are aware of the safe distance requirements and are trained to these manual updates. (4) Dispose of obsolete guides. Technical support number are 1-800-348-9011 or 1-978-421-9460 and are available to assist users with any aspect of this notice
Quantity in commerce
2361 units
Distribution
Worldwide distribution - US Nationwide and the countries of United Arab Emirates, Argentina, Australia, Azerbaijan, Bangladesh, Belgium, Bahrain, Bolivia, Brazil, Canada, Switzerland, China, Colombia, Czech Republic, Germany, Spain, Great Britain (UK), Greece, Guatemala, Guam, Hong Kong, Croatia, Hungary, Indonesia, Israel, India, Iraq, Italy, Japan, South Korea, Lebanon, Myanmar, Mexico, Malaysia, Netherlands, New Zealand, Oman, Panama, Philippines, Pakistan, Poland, Puerto Rico, Portugal, Qatar, Romania, Saudi Arabia, Singapore, Thailand, Turkey, Taiwan, Vietnam.
Product codes and lots
Part Numbers/UDI Zoll 731 Defibrillators: 1. 799-EGL2-02 00847946020910 2. 799-EGL2-02-01 00847946023454 3. 799-EGL2-02-04 00847946036867 4. 799-EGL2-02-05 00847946036874 5. 799-EGL2-02-08 6. 799-EGL2-02-10 00847946036881 7. 799-EGL2-02-11 8. 799-EGL2-02-15 00847946050665 9. 799-EGL2-02-16 10. 799-EMBP-05-01 00847946024185 11. 799-EMBP-05-02 12. 799-EMBP-05-08 13. 799-EMBP-05-10 14. 799-EMBP-05-11 15. 799-EMBP-05-12 00847946036959 16. 799-EMBP-05-12-68 17. 799-EMBP-05-16 18. 799-EMBP-05-23 19. 799-EMVP-05 00847946020767 20. 799-EMVP-05-01 00847946023430 21. 799-EMVP-05-05 00847946034351 22. 799-EMVP-05-10 23 799-EMVP-05-15 00847946049850 24 799-EMVP-05-16 Manual Part Numbers 906-0731-01; 9650-002365-01 (US); 906-0731-01-05 (CAN); 906-0731-01-10; 9650-0002363-01; 906-0731-01 (Commercial); 906-0731-03 (Military); 9650-0002363-01; 9650-0002363-05; 906-0731-01; 906-0731-01-05 (CAN); 906-0731-01-04; 906-0731-01-05; 906-0731-01-08; 906-0731-01-10; 906-0731-01-11; 906-0731-01-15; 906-0731-01-16; 9650-0002363-01; 9650-0002363-02; 9650-0002363-08; 9650-0002363-10; 9650-0002363-11; 9650-0002363-12; 9650-0002363-16; 9650-0002363-23; 906-0731-01 (Commercial); 906-0731-03 (Military); 9650-0002363-01; 9650-0002363-05; 9650-0002363-10; 9650-0002363-15; 9650-0002363-16 Quick Reference Guide (QRG) Part Numbers: 907-0731-04; 9652-000499-01; 9652-000511-01; 907-0731-04; 9652-000511-01; 907-0731-04; 9652-000499-01; 9652-000499-04; 9652-000499-08; 9652-000499-10; 9652-000499-11; 907-0731-04; 9652-000499-16; 9652-000511-01; 9652-000511-02; 9652-000511-08; 9652-000511-10; 9652-000511-11; 9652-000511-12; 9652-000511-16; 9652-000511-23; 907-0731-04; 9652-000511-01; 9652-000511-05; 9652-000511-10; 9652-000511-15; 9652-000511-16

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2130-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2130-2024Class IOngoingVoluntary: Firm initiated