Z-2029-2024Class II
OES 4000 Hysteroscope, Model/Catalog Number: A4674A
OLYMPUS Corporation of the Americas / initiated 2024-04-18 / Open, Classified
Olympus Corporation of the Americas began a recall of OES 4000 Hysteroscope, Model/Catalog Number: A4674A on . FDA classified it as Class II and gives the reason as "The required leakage test was not performed following of a re-work for a non-conformity. This could potentially result in liquid ingress into the eyepiece and cause a foggy image". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-2029-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- OES 4000 Hysteroscope, Model/Catalog Number: A4674A
- Product code
- HIH Hysteroscope (And Accessories)
- Reason for recall
- The required leakage test was not performed following of a re-work for a non-conformity. This could potentially result in liquid ingress into the eyepiece and cause a foggy image.
- FDA root cause
- Process change control
- Action
- On May 9, 2024, URGENT: MEDICAL DEVICE REMOVAL letters were sent to the consignee. Actions Required: Our records indicate that your facility has purchased one or more of the affected products. Therefore, Olympus requires you to take the following actions: 1. Carefully read the content of this notification. 2. Examine your inventory and identify the above listed device with serial number. 3. Please contact Customer Service at at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus, and we will replace this device at no charge. 4. Olympus requests that you acknowledge receipt of this letter through the a. Go to https://olympusamerica.com/recall b. Enter the recall number "XXXX . c. Complete the form as instructed and include your account ID number: XXXXXXXX 5. If you have further distributed this product, identify your customers, and forward them this notification. Olympus requests that you report any complaints, including foggy image, to the Technical Assistance Center (TAC) at 1- 800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me directly by phone at (647) 999-3203 or by e-mail at [email protected]
- Quantity in commerce
- 1 unit
- Distribution
- US Nationwide distribution in the state of NY.
- Product codes and lots
- Model/Catalog Number: A4674A, UDI-DI: 04042761006361, Serial Number: 805904;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2029-2024 | Class II | Ongoing | Voluntary: Firm initiated |