Z-2257-2024Class II

Neurostimulation Kit (Ported, Dual 8, 40 cm), REF: 71005-US Peripheral Nerve Stimulation Kit (Ported, Dual 8...

Nalu Medical / initiated 2024-06-03 / Open, Classified / root cause: manufacturing process

Nalu Medical, Inc. began a recall of Neurostimulation Kit (Ported, Dual 8, 40 cm), REF: 71005-US Peripheral Nerve Stimulation Kit (Ported, Dual 8, 40 cm), REF: 71019-US Components of the Nalu... on . FDA classified it as Class II and gives the reason as "Neurostimulation Systems have a manufacturing non-conformance that may reduce the efficiency of the radiofrequency signals between the Implantable Pulse Generator (IPG) and the Therapy Disc, causing the IPG to be...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2257-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Nalu Medical
Product
Neurostimulation Kit (Ported, Dual 8, 40 cm), REF: 71005-US Peripheral Nerve Stimulation Kit (Ported, Dual 8, 40 cm), REF: 71019-US Components of the Nalu Spinal Cord Stimulation System.
Product code
GZF Stimulator, Peripheral Nerve, Implanted (Pain Relief)
Reason for recall
Neurostimulation Systems have a manufacturing non-conformance that may reduce the efficiency of the radiofrequency signals between the Implantable Pulse Generator (IPG) and the Therapy Disc, causing the IPG to be non-functional, which causes the device to be non-functional so it will no longer inhibit the transmission of pain signals to the brain.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On 6/3/24, recall notices were mailed and emailed to Physicians managing patients with Implanted Pulse Generators (IPG) who were informed of the following: 1) Four of the IPGs have been found to have low RF efficiency readings. 2) Firm recommends monitoring the patients with IPGs with low RF efficiency readings through healthcare visits that occur to ensure proper functioning of the Neurostimulation Kit/Peripheral Nerve Stimulation Kit and working with your field representative to address any issues in the event the device is not performing as intended since there are other factors that could cause similar product behaviors. If troubleshooting is unable to restore functionality, it is recommended that you make a clinical assessment around replacement surgery. Ultimately, the decision to act and/or communication with the patient will be at your discretion. 3) If you find a device that has been identified as part of the recalled product to be nonfunctional, you should notify us immediately through your local sales representative, our technical support line at (800) 618-3402 4) Complete and return the response form via email to [email protected] 5) If you have any questions, please call your local sales representative at (480) 313-9745.
Quantity in commerce
13
Distribution
US Nationwide distribution in the states of MO, IL, CA, TX, TN, OK, AZ, AR.
Product codes and lots
UDI-DI: 71005-US: 00812537036137, 00812537033600, 00812537033518, 00812537030975, 00812537030524. 71019-US: 00812537036441, 00812537035604, 00812537035598 Serial Numbers: H01377, H01382, H01432, H01419, H01387, H01445, H01415, H01441, H01373, H01428, H01378, H01438, H01431

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2257-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2257-2024Class IIOngoingVoluntary: Firm initiated