Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2632-2024Class II

Medline procedure packs containing Shenli syringes, labeled as follows: 1) NEURO ANGIO PACK, Part Number...

MEDLINE INDUSTRIES, LP - Northfield / initiated 2024-04-08 / Open, Classified / root cause: manufacturing process

MEDLINE INDUSTRIES, LP - Northfield began a recall of Medline procedure packs containing Shenli syringes, labeled as follows: 1) NEURO ANGIO PACK, Part Number DYNDA1431A; 2) CVC INSERT BUNDLE 16CM, Part Number... on . FDA classified it as Class II and gives the reason as "Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2632-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Medline procedure packs containing Shenli syringes, labeled as follows: 1) NEURO ANGIO PACK, Part Number DYNDA1431A; 2) CVC INSERT BUNDLE 16CM, Part Number ECVC3260; 3) CENTRAL LINE INSERTION KIT, Part Number ECVC7440; 4) CENTRAL LINE INSERTION KIT, Part Number ECVC7445; 5) CENTRAL LINE INSERTION KIT W/20CM CATH, Part Number ECVC7445A; 6) VANTEX 7FR 3L 20CM CVC BUNDLE US, Part Number ECVC8045; 7) CVC DOUBLE LUMEN 16CM, Part Number ECVC8105
Product code
LRP Tray, Surgical
Reason for recall
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request these syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.
Quantity in commerce
1,679,067 units in total
Distribution
Worldwide distribution.
Product codes and lots
DYNDA1431A, Lot Number 2019021190; ECVC3260, Lot Number 2019091850; ECVC3260, Lot Number 2019120950; ECVC7440, Lot Number 2021071950; ECVC7440, Lot Number 2021092150; ECVC7445, Lot Number 2021052850; ECVC7445, Lot Number 2021062250; ECVC7445A, Lot Number 2022012850; ECVC7445A, Lot Number 2023010390; ECVC8045, Lot Number 2023031590; ECVC8105, Lot Number 2023112190

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2632-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2623-2024Class IIOngoingVoluntary: Firm initiated
Z-2624-2024Class IIOngoingVoluntary: Firm initiated
Z-2625-2024Class IIOngoingVoluntary: Firm initiated
Z-2626-2024Class IIOngoingVoluntary: Firm initiated
Z-2627-2024Class IIOngoingVoluntary: Firm initiated
Z-2628-2024Class IIOngoingVoluntary: Firm initiated
Z-2629-2024Class IIOngoingVoluntary: Firm initiated
Z-2630-2024Class IIOngoingVoluntary: Firm initiated
Z-2631-2024Class IIOngoingVoluntary: Firm initiated
Z-2632-2024Class IIOngoingVoluntary: Firm initiated
Z-2633-2024Class IIOngoingVoluntary: Firm initiated
Z-2634-2024Class IIOngoingVoluntary: Firm initiated
Z-2635-2024Class IIOngoingVoluntary: Firm initiated