CareFusion 303, Inc. began a recall of BD Pyxis: MedStation ES (Main), REF: 323; MedStation ES Tower, REF: 352; Anesthesia Station ES, REF: 327; CII Safe ES, REF: 1116-00; CII Safe ES, Desktop PC... on . FDA classified it as Class II and gives the reason as "If automated dispensing cabinets have specific software versions, and Component Manager was configured to "installed mode", and software patch KB 5033688 was installed in the production environment, then this will will...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2374-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- CareFusion 303, Inc
- Product
- BD Pyxis: MedStation ES (Main), REF: 323; MedStation ES Tower, REF: 352; Anesthesia Station ES, REF: 327; CII Safe ES, REF: 1116-00; CII Safe ES, Desktop PC, Wired Scanner, REF: 107-254-01; CII Safe ES, Desktop PC, Wireless Scanner, REF: 107-255-01; CII Safe ES, Desktop PC, Wired Scanner 1.7.3, REF: 107-256-01; CII Safe ES, Desktop PC, Wireless Scanner 1.7.3, REF: 107-257-01
- Product code
- BRY Cabinet, Table And Tray, Anesthesia
- Reason for recall
- If automated dispensing cabinets have specific software versions, and Component Manager was configured to "installed mode", and software patch KB 5033688 was installed in the production environment, then this will will lead to an error that prevents the user from accessing the dispensing software application, which may delay accessing medication.
- Root cause tag
- software
- FDA root cause
- Software change control
- Action
- On 6/6/24, correction notices were distributed to customers who were informed that their devices had been remediated. That they had been contacted by phone in May 2024 to schedule device remediation. The firm re-configured the component manager from "installed mode" to "disabled mode" and checked if the Microsoft patch was present on each device. If the patch was identified on a device, it was removed. All devices were then configured to "managed mode" and rebooted to enable the system appropriately. Customers are asked to complete and return the response form to Email [email protected] For safety information about the system not covered in this document, call the firm's Technical Support Center at 1-800-727-6102, Phone hours: 24 hours, 7 days a week. Additional contacts include: 1) Regional Complaint Center, 1-844-823-5433 Say "Product Complaints" when prompted Mon-Fri 8:00am and 5:00pm CT, Email: [email protected] 2) Post Market Quality Field Action Operations, [email protected]
- Quantity in commerce
- 326
- Distribution
- Worldwide - US Nationwide distribution including in the states of OR, KS, NE, LA, ND, OK, AL, AK, WI, WV, WA, NV, PR, TX, CO, OH, CA, MO, IA, AZ, IN, AR, IL, MN, HI, NM, FL, SD, UT, NY, VA, GA, MA, NC and the countries of TW, CA.
- Product codes and lots
- REF/UDI-DI: 323/10885403512667, 352/10885403512674, 327/10885403477836, 1116-00/10885403512605 Software Version: ES 1.7.x Serial Numbers: 13534745, 16220738, 16446299, 16446303, 16446302, 16247443, 16247440, 15962990, 16417275, 16446297, 16446296, 16242783, 16234266, 15425093, 16446295, 16446291, 16242784, 16417277, 15644176, 14019332, 16212329, 16221517, 16395857, 16059358, 16254191, 15319466, 15154945, 15153602, 16215325, 15154942, 14959051, 16395855, 16395853, 16215313, 16361893, 14826969, 14826972, 14826971, 14826973, 14232856, 16241342, 13877154, 15148881, 15148860, 15148880, 15148882, 15854601, 15148883, 15148865, 15148885, 15160417, 16212961, 15854595, 15148859, 15148862, 15148861, 15148863, 15148866, 15148868, 15854596, 15854600, 15148870, 15854597, 15854598, 15854599, 15148873, 15148864, 15148867, 15148884,15148878, 15148879, 16324054, 16223557, 15136666, 13841400, 14138516, 15519749, 16446315, 16141586, 16416617, 16416618, 16416622, 16416619, 16416621, 16416623, 16416616, 16416615, 13781497, 14557551, 16212312, 15500549, 14700155, 16212346, 16214431, 16214432, 16214430, 16214429, 16212344, 16247441, 16247444, 16320157, 16335215, 16204472, 15154938, 14813899, 14697245, 15642731, 16209866, 15135154, 16214435, 16446294, 16227849, 16446311, 16010961, 16375509, 15303370, 16415510, 14668921, 16239128, 16346806, 16242800, 16204479, 16352591, 16324052, 16222213, 16212322, 16271477, 15015227, 16214427, 16373774, 15978927, 16358533, 16215391, 16446313, 16446308, 16247439, 16247445, 16247442, 16347745, 16347746, 14157588, 16271458, 16346807, 15196235, 16322403, 16352593, 14728645, 16212338, 15519703, 16217225, 14806556, 16352598, 16352590, 16352595, 16352596, 15263493, 16216425, 16416639, 16416102, 16416104, 16416632, 16416108, 16416637, 16416110, 16416111, 16416106, 16416638, 14987500, 16234192, 16405456, 16405459, 15136661, 16242578, 16271457, 15911127, 15954728, 16271463, 16271465, 14966741, 16214437, 16324055, 16225327, 16407028, 14613489, 15997584, 14442210, 16357615, 16357616, 16307328, 13839287, 13839299, 15136665, 15045814, 16054085, 15642724, 16244068, 16346335, 16283945, 16231509, 16277495, 16240201, 16211337, 16291566, 16212335, 15933030, 16242781, 16346808, 15135153, 16204480, 16271460, 16212330, 16324056, 16215395, 16214434, 15454500, 16173369, 16423187, 15895362, 16242797, 16204482, 16274624, 16223058, 16223059, 16274625, 15540317, 16223060, 16274626, 16257759, 16352588, 16446301, 12345678, 16271461, 16352594, 16212342, 16120462, 16235542, 16242799, 16212336, 16468210, 16195795, 16211170, 15570657, 16223556, 16412617, 16412618, 16412615, 16412616, 16321316, 16022041, 16242801, 16267770, 15136667, 15136668, 16174009, 14956271, 15236068, 15236067, 16218170, 16061265, 14572826, 16128483, 16128482, 16143915, 15398129, 15449307, 16174023, 16174021, 16143913, 16174036, 16174008, 16174019, 16174020, 15449302, 15691588, 15449303, 16227720, 16446290, 16242780, 16242802, 16405462, 16116814, 16212339, 16008937, 16215393, 16215389, 16215398, 16223558, 15147150, 16342242, 16253003, 16253008, 15495195, 16149377, 16094701, 15560137, 16296089, 16296090, 16296091, 16296092, 13347728, 14826153, 16417216, 16417243, 16417242, 16324044, 16415642, 16419185, 16415641, 16415639, 16415640, 16242803, 16242804, 16212319, 15136664, 16262416, 15136663, 16317160, 16212343, 15136670
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2374-2024 | Class II | Ongoing | Voluntary: Firm initiated |