Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2212-2024Class II

Product not distributed in the US. Implantable Cardioverter Defibrillators: a. VIVA QUAD XT CRT-D, Model...

Medtronic, Inc. / initiated 2024-04-16 / Open, Classified / root cause: manufacturing process

Medtronic Inc. began a recall of Product not distributed in the US. Implantable Cardioverter Defibrillators: a. VIVA QUAD XT CRT-D, Model Number DTBA2QQ. b. CLARIA MRI CRT-D SureScan, Model... on . FDA classified it as Class II and gives the reason as "Potential for manufacturing defect. Weld crack led to devices failing at the leak check step in manufacturing". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2212-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
Product not distributed in the US. Implantable Cardioverter Defibrillators: a. VIVA QUAD XT CRT-D, Model Number DTBA2QQ. b. CLARIA MRI CRT-D SureScan, Model Number DTMA2D1. c. CLARIA MRI QUAD CRT-D SureScan, Model Number DTMA2QQ. d. AMPLIA MRI CRT-D SureScan, Model Number DTMB2D1. e. AMPLIA MRI CRT-D SureScan, Model Number DTMB2D4. f. AMPLIA MRI QUAD CRT-D SureScan, Model Number DTMB2Q1. g. AMPLIA MRI QUAD CRT-D SureScan, Model Number DTMB2QQ. h. COMPIA MRI QUAD CRT-D SureScan, Model Number DTMC2QQ. i. EVERA MRI XT DR SureScan, Model Number DDMB2D4. j. VISIA AF MRI XT VR SureScan, Model Number DVFB2D1. k. VISIA AF MRI S VR SureScan, Model Number DVFC3D4. l. VISIA AF MRI XT VR SureScan, Model Number DVFB2D4.
Product code
LWS Implantable Cardioverter Defibrillator (Non-Crt)
Reason for recall
Potential for manufacturing defect. Weld crack led to devices failing at the leak check step in manufacturing.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On about 04/18/2024, Medtronic notified customers that they were retrieving distributed units that may have undergone a specific manufacturing sequence that requires additional engineering evaluation. They requested that customers acknowledge the letter with signature.
Quantity in commerce
82 units
Distribution
Worldwide - US Nationwide distribution in the states of CA, CT, FL, IN, MA, MI, NC, NH, NJ, NM, NY, OH, TX, VA and the countries of Australia, Austria, Bulgaria, Canada, China, Croatia, Czech Republic, France, Georgia, Germany, Greece, Guadeloupe, Iceland, Ireland, Italy, Jordan, Kosovo, Latvia, Moldova, Republic Of, Netherlands, Northern Ireland, Norway, Oman, Panama, Poland, Portugal, Qatar, Reunion, Slovakia, Slovenia, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.
Product codes and lots
a. GTIN: 00763000612269, Serial Numbers: BLG628370S, BLG628372S, BLG628373S. b. GTIN: 00763000206284, Serial Number: RPO606047S. c. GTIN: 00763000206222, Serial Numbers: RPB629895S, RPB629900S, RPB629902S, RPB629906S, RPB629909S, RPB629912S. d. GTIN: 00763000206307, Serial Number: RPQ610208S. e. GTIN: 00763000206291, Serial Numbers: RPK614721S, RPK614728S. f. GTIN: 00763000206253, Serial Numbers: RPP607499S, RPP607504S, RPP607517S, RPP607518S, RPP607522S, RPP607586S. g. GTIN: 00763000206246, Serial Numbers: RPG643295S, RPG643318S; GTIN 00763000611835, Serial Numbers: RPG643198S, RPG643199S, RPG643200S, RPG643203S, RPG643206S, RPG643209S, RPG643211S, RPG643213S, RPG643218S, RPG643219S, RPG643220S, RPG643221S, RPG643231S, RPG643251S. h. GTIN: 00763000206260, Serial Numbers: RPF648668S, RPF648672S, RPF648674S, RPF648675S, RPF648676S, RPF648677S, RPF648681S, RPF648684S, RPF648685S, RPF648686S, RPF648691S, RPF648693S, RPF648702S, RPF648758S, RPF648760S, RPF648761S, RPF648762S, RPF648763S, RPF648768S, RPF648769S, RPF648770S, RPF648771S, RPF648844S, RPF648846S, RPF648847S. i. GTIN: 00763000206147, Serial Numbers: PGZ663023S, PGZ663203S, PGZ663208S, PGZ663211S, PGZ663216S, PGZ663219S, PGZ663241S, PGZ663264S. j. GTIN: 00763000612191, Serial Numbers: CWH606546S, CWH606548S, CWH606551S, CWH606552S, CWH606553S. k. GTIN: 00763000206352, Serial Numbers: PMX622050S, PMX622052S, PMX622055S, PMX622056S, PMX622057S, PMX622059S. l. GTIN: 00763000206338, Serial Numbers: PLX622054S, PLX622056S, PLX622072S.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2212-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2190-2024Class IIOngoingVoluntary: Firm initiated
Z-2191-2024Class IIOngoingVoluntary: Firm initiated
Z-2192-2024Class IIOngoingVoluntary: Firm initiated
Z-2193-2024Class IIOngoingVoluntary: Firm initiated
Z-2194-2024Class IIOngoingVoluntary: Firm initiated
Z-2195-2024Class IIOngoingVoluntary: Firm initiated
Z-2196-2024Class IIOngoingVoluntary: Firm initiated
Z-2197-2024Class IIOngoingVoluntary: Firm initiated
Z-2198-2024Class IIOngoingVoluntary: Firm initiated
Z-2199-2024Class IIOngoingVoluntary: Firm initiated
Z-2200-2024Class IIOngoingVoluntary: Firm initiated
Z-2201-2024Class IIOngoingVoluntary: Firm initiated
Z-2202-2024Class IIOngoingVoluntary: Firm initiated
Z-2203-2024Class IIOngoingVoluntary: Firm initiated
Z-2204-2024Class IIOngoingVoluntary: Firm initiated
Z-2205-2024Class IIOngoingVoluntary: Firm initiated
Z-2206-2024Class IIOngoingVoluntary: Firm initiated
Z-2207-2024Class IIOngoingVoluntary: Firm initiated
Z-2208-2024Class IIOngoingVoluntary: Firm initiated
Z-2209-2024Class IIOngoingVoluntary: Firm initiated
Z-2210-2024Class IIOngoingVoluntary: Firm initiated
Z-2211-2024Class IIOngoingVoluntary: Firm initiated
Z-2212-2024Class IIOngoingVoluntary: Firm initiated