Volara System. Single Patient Use (SPU) Circuit 5kit for Volara and Single Patient Use (SPU) Circuit kit for...
Baxter Healthcare Corporation began a recall of Volara System. Single Patient Use (SPU) Circuit 5kit for Volara and Single Patient Use (SPU) Circuit kit for Volara. Intermittent positive pressure breathing... on . FDA classified it as Class I and gives the reason as "Potential of the handset plug to disconnect from the nebulizer port on the blue ventilator adapter. The handset plug is required to ensure proper operation and ventilator gas flow". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2321-2024
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Volara System. Single Patient Use (SPU) Circuit 5kit for Volara and Single Patient Use (SPU) Circuit kit for Volara. Intermittent positive pressure breathing device.
- Product code
- NHJ Device, Positive Pressure Breathing, Intermittent
- Reason for recall
- Potential of the handset plug to disconnect from the nebulizer port on the blue ventilator adapter. The handset plug is required to ensure proper operation and ventilator gas flow.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Baxter Healthcare notified customers on about 05/30/2024 via letter sent first-class mail. The letter instructed facilities to immediately locate, isolate, and cease all use of the affected lot numbers, return affected units on hand, and complete and return the acknowledgement form. If the product was further distributed, they were instructed to notify customers if product was further distributed. The letter instructed patients to immediately locate, isolate, and cease all use of the affected lot numbers, return affected units on hand, however if they do not have other options in stock, in order to not delay therapy, you may continue to use with caution. Visually inspect the nebulizer port prior to therapy and assess for ventilator gas flow leaks for the duration of therapy. They were also instructed to complete and return the acknowledgement form.
- Quantity in commerce
- 58,220 units
- Distribution
- Nationwide distribution. International distribution to Canada, France, and Germany.
- Product codes and lots
- Product Codes: M08473/M08474; Lot numbers: 2171WA, 2237WA, 2238WA, 2241WA, 2242WA, 2243WA, 2244WA, 2249WA, 2250WA, 2251WA, 2285WA, 2286WA, 2290WA, 2297WA, 2298WA, 2299WA, 2307WA, 2308WA, 2311WA, 2332WA, 2333WA, 2341WA, 2342WA, 2349WA, 2355WA, 2356WA, 2357WA, 2363WA, 2364WA, 3019WA, 3061WA, 3067WA, 3068WA, 3069WA, 3072WA, 3073WA, 3074WA, 3080WA, 3081WA, 3094WA, 3095WA, 3096WA, 3100WA, 3101WA, 3102WA, 3103WA, 3109WA, 3110WA, 3111WA, 3123WA, 3124WA, 3152WA, 3156WA, 3157WA, 3158WA, 3213WA, 3214WA, 3215WA, 3219WA, 3240WA, 3241WA, 3268WA, 3270WA, 3271WA, 3272WA, 3297WA, 3298WA, 3299WA, 3300WA, 3303WA, 3304WA, 3305WA, 3306WA, 3310WA, 3311WA, 3313WA, 3326WA, 3327WA, 3328WA, 3331WA.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2321-2024 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2322-2024 | Class I | Ongoing | Voluntary: Firm initiated |