Z-2179-2024Class II

RingLoc Bi-Polar Hip System Acetabular Cup, 41 mm Outer Diameter - Used in hip hemi-arthroplasty Item Number...

Zimmer GmbH / initiated 2024-05-17 / Open, Classified / root cause: labeling and UDI

Biomet, Inc. began a recall of RingLoc Bi-Polar Hip System Acetabular Cup, 41 mm Outer Diameter - Used in hip hemi-arthroplasty Item Number: 11-165206 on . FDA classified it as Class II and gives the reason as "41 mm acetabular outer heads were potentially assembled with a 42 mm locking ring, may lead to joint instability leading requiring surgical intervention". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2179-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Zimmer GmbH
Product
RingLoc Bi-Polar Hip System Acetabular Cup, 41 mm Outer Diameter - Used in hip hemi-arthroplasty Item Number: 11-165206
Product code
JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
Reason for recall
41 mm acetabular outer heads were potentially assembled with a 42 mm locking ring, may lead to joint instability leading requiring surgical intervention
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
Zimmer Biomet issued Urgent Medical Device Recall Letter on 5/17/24 to Distributors and End-Users via FedEx and email. Letter states reason for recall, health risk and action to take: Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3.If the product has been further distributed, provide your customers with the recall notice and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 5. If you have further questions or concerns after reviewing this notice, please call custom
Quantity in commerce
12 units
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, NJ, PA, TX, VA and the country of Chile.
Product codes and lots
Lot Number / UDI Number (1) 66540443 (01)00880304001923(17)290408(10)66540443; (2) 66601057 (01)00880304001923(17)290405(10)66601057; (3) 66623486 (01)00880304001923(17)290327(10)66623486; (4) 66635486 (01)00880304001923(17)290327(10)66635486

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2179-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2179-2024Class IIOngoingVoluntary: Firm initiated