Z-2247-2024Class II

First Source iQ Flex M. Portable cart and platform that supports an X-ray system and laptop computer

First Source Inc / initiated 2024-05-09 / Open, Classified / root cause: design

First Source Inc began a recall of First Source iQ Flex M. Portable cart and platform that supports an X-ray system and laptop computer on . FDA classified it as Class II and gives the reason as "Potential for the bolt that secures the positioning arm to the gas spring to come loose and fall onto patient". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-2247-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
First Source Inc
Product
First Source iQ Flex M. Portable cart and platform that supports an X-ray system and laptop computer.
Product code
IZL System, X-Ray, Mobile
Reason for recall
Potential for the bolt that secures the positioning arm to the gas spring to come loose and fall onto patient.
Root cause tag
design
FDA root cause
Device Design
Action
Consignees were emailed an URGENT: MEDICAL DEVICE RECALL notification, dated 5/29/24. The notice instructs consignees to follow the recommended preventive maintenance schedule and visually check the device before use to identify if the affected bolt is coming loose. The recalling firm is currently testing a spring clip which will keep the bolt from falling out and will send this part to customers at no charge once approved. Consignees are asked to return the provided response form. Consignees with any questions can contact the firm by email at [email protected] or by phone at 1-800-349-5980 (option 2) from 9:00 AM to 5:00 PM EST from Monday through Friday.
Quantity in commerce
137 units
Distribution
Worldwide - US Nationwide distribution in the states of AR, CA, FL, GA, IL, IN, LA, MA, MD, MI, MO, MS, NC, NY, OK, PA, TN, TX, & VA. The countries of Korea & Serbia.
Product codes and lots
Model No. Pending; UDI-DI: G221500MC0740; Serial No. 4001, 4002, 4003, 4004, 4005, 4006, 4007, 4008, 4009, 4010, 4011, 4012, 4013, 4014, 4015, 4016, 4017, 4018, 4019, 4020, 4021, 4022, 4023, 4024, 4025, 4026, 4027, 4028, 4029, 4030, 4031, 4032, 4033, 4034, 4035, 4036, 4037, 4038, 4039, 4040, 4041, 4042, 4043, 4044, 4045, 4046, 4047, 4048, 4049, 4050, 4051, 4052, 4053, 4054, 4055, 4056, 4057, 4058, 4059, 4060.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2247-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2246-2024Class IIOngoingVoluntary: Firm initiated
Z-2247-2024Class IIOngoingVoluntary: Firm initiated