Z-2529-2024Class II

Leica HistoCore Arcadia H, REF: 14039354100 and 14039357258, Paraffin Embedding Station

Leica Biosystems Nussloch GmbH / initiated 2024-05-10 / Open, Classified / root cause: design

LEICA BIOSYSTEMS NUSSLOCH GMBH began a recall of Leica HistoCore Arcadia H, REF: 14039354100 and 14039357258, Paraffin Embedding Station on . FDA classified it as Class II and gives the reason as "An issue with safe usage of the device was identified whereby toxic smoke and internal fire developed resulting from a problem associated with the sealing of the paraffin tank". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-2529-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Leica Biosystems Nussloch GmbH
Product
Leica HistoCore Arcadia H, REF: 14039354100 and 14039357258, Paraffin Embedding Station.
Product code
IDW Dispensers, Paraffin
Reason for recall
An issue with safe usage of the device was identified whereby toxic smoke and internal fire developed resulting from a problem associated with the sealing of the paraffin tank.
Root cause tag
design
FDA root cause
Device Design
Action
The recalling firm issued letters dated 4/30/2024 via email on 5/10/2024. The letter provided the model and serial number of the devices involved, a description of the problem, and immediate actions to be taken. The consignee is to check to see if they have one of the affected instruments and determine whether the Arcadia H is in active use in the facility. If it is, the consignee is to discontinue the use of it immediately and disconnect it from main voltage. If it has not been decommissioned at the time of receipt of the letter, the consignee is to reach out to their sales representative. The sales representative will work with the consignee to determine the disposition of the device according to their local regulations. The consignee is requested to pass the notification letter to users of the device and to all within their organization who need to know. The consignee is to confirm receipt of the letter within 5 days by completing the attached acknowledgement form and returning a scanned copy via email.
Quantity in commerce
180 devices
Distribution
US Nationwide, including government distribution.
Product codes and lots
Model/Catalog Number: 14039354100; UDI-DI: 04049188194763, Serial numbers: 58, 62, 98, 99, 101, 117,120, 121, 122, 123, 124, 125, 127, 128, 129, 177, 178, 179, 180, 181, 185, 186, 218, 222, 223, 279, 282, 284, 286, 287, 288, 289, 290, 291, 292, 314, 315, 317, 318, 319, 321, 322, 353, 354, 355, 356, 357, 358, 360, 363, 365, 366, 371, 478, 480, 483, 484, 487, 488, 489, 490, 491, 492, 553, 560, 561, 562, 584, 586, 588, 589, 590, 593, 597, 643, 644, 647, 649, 650, 651, 652, 653, 656, 658, 659, 697, 698, 705, 710, 714, 715, 716, 717, 739, 740, 741, 742, 743, 757, 819, 820, 821, 822, 825, 826, 827, 828, 829, 830, 831, 832, 878, 879, 880, 881, 882, 883, 884, 885, 886, 887, 888, 889, 890, 891, 892, 893, 894, 896, 897, 898, 900, 901, 904, 906, 907, 909, 911, 919, 921, 922, 923, 924, 927, 928, 930, 1113, 1114, 1115, 1116, 1117, 1118, 1123, 1124, 1198, 1199, 1200, 1201, 1202, 1203, 1204, 1205, 1206, 1207, 1212, 1216, 1217, 1266, 1267, 1268, 1271, 1273, 1274, 1277, 1278, 1279, 1280, 1283, and 1285. Model/Catalog Number: 14039357258; UDI-DI: 04049188205971, serial number 5033.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2529-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2529-2024Class IIOngoingVoluntary: Firm initiated