Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2428-2024Class II

Beckman Coulter, REF: OSR61173, IGM 4x14 mL R1, 4x11 mL R2

Siemens Healthcare Diagnostics Inc. / initiated 2024-05-27 / Open, Classified / root cause: manufacturing process

Beckman Coulter Inc. began a recall of Beckman Coulter, REF: OSR61173, IGM 4x14 mL R1, 4x11 mL R2 on . FDA classified it as Class II and gives the reason as "LIH Influence check settings for Lipemia are incorrect which may result patient samples being reported as false low results or cause a high result to report as normal". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2428-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Beckman Coulter, REF: OSR61173, IGM 4x14 mL R1, 4x11 mL R2
Product code
CFN Method, Nephelometric, Immunoglobulins (G, A, M)
Reason for recall
LIH Influence check settings for Lipemia are incorrect which may result patient samples being reported as false low results or cause a high result to report as normal
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On May 23, 2024, Beckman Coulter issued a "Urgent Medical Device Recall" notification via: xxx. 1. Update the LIH Influence Check settings for Lipemia on AU480, AU680, DxC 500 AU and DxC 700 AU analyzers if they are enabled for IgM. 2. To determine if the LIH Influence Check is enabled for IgM and to update your Lipemia setting, see customer communication. 3. Beckman Coulter recommends sharing the content of this letter with your laboratory and/or Medical Director to determine if a review of previous patient test results should be conducted. . Discontinuance or disposal of this product is not necessary. . Retain a copy of this letter as it serves as current labelling. . Per the IFU, avoid highly Lipemia samples when using the IgM assay. 4. Please complete the response form within 10 days in one of the following ways: - Electronically, if you received this communication via email./Manually, complete and return the enclosed Response Form. 5. If you have any questions regarding this notice, please contact our Customer Support Center: From our website: http://www.beckmancoulter.com/By phone: call 1-800-854-3633 in the United States. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Quantity in commerce
13,565 units
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, WA, WI, WV, WY and the countries of Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China (DX), Colombia, Croatia, Cyprus, Czechia, Denmark, Ecuador, Egypt, Estonia, France, Georgia, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Iraq, Ireland, Israel, Italy, Jordan, Kazakhstan, Korea, Republic of, Lebanon, Lithuania, Malaysia, Mayotte, Mexico, Morocco, Mozambique, Netherlands, New Zealand, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Province of China, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, United States of America, Viet Nam, Yemen.
Product codes and lots
UDI: 15099590011598/ All Lots: 2548/Exp. 2024-06-01, 2549/Exp.2024-10-01, 2550/Exp. 2024-11-01, 2551/Exp. 2025-04-01, 2552/Exp. 2025-07-01, 2553/Exp. 2025-10-01, 2554/ Exp. 2026-03-01

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2428-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2428-2024Class IIOngoingVoluntary: Firm initiated