Z-2367-2024Class II

Abbott Proclaim DRG Implantable Pulse Generator, REF 3664, Sterile EO, Rx Only

Abbott Medical / initiated 2024-05-16 / Open, Classified / root cause: supplier

Abbott Medical began a recall of Abbott Proclaim DRG Implantable Pulse Generator, REF 3664, Sterile EO, Rx Only on . FDA classified it as Class II and gives the reason as "The duration between the implantable pulse generator reaching the elective replacement indicator threshold and end of service may be shorter than indicated in product labeling". As of , the recall status on record is Open, Classified, and MDM tags the root cause as supplier.

Recall record

Recall number
Z-2367-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Abbott Medical
Product
Abbott Proclaim DRG Implantable Pulse Generator, REF 3664, Sterile EO, Rx Only
Product code
PMP Dorsal Root Ganglion Stimulator For Pain Relief
Reason for recall
The duration between the implantable pulse generator reaching the elective replacement indicator threshold and end of service may be shorter than indicated in product labeling.
Root cause tag
supplier
FDA root cause
Vendor change control
Action
On June 6, 2024, Abbott issued a Urgent Medical Device Correction notification to affected consignees. Abbott asked consignees to take the following actions: 1. This issue will continue to affect your patients for the foreseeable future. Abbott is working on an update to align the product labeling with actual device performance. 2. When an ERI notification appears, see the following considerations to make a determination with your patients on timing of IPG replacement within the customer notification. 3. See the customer notification for recommended strategies for preventing lapse in therapy. 4. Please return a completed Acknowledgement Form and maintain a record of this notice along with a copy of the completed Acknowledgement Form. 5. For questions about this issue or reported complaints, please contact your Abbott representative, or Abbott Technical Support at 1-800-727-7846. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Quantity in commerce
21,843 units
Distribution
Worldwide - US Nationwide distribution including in the states of ALABAMA, ALASKA, ARIZONA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DELAWARE, DISTRICT OF COLUMBIA, FLORIDA, GEORGIA, GUAM, HAWAII, IDAHO, ILLINOIS, INDIANA, IOWA, KANSAS, KENTUCKY, LOUISIANA,MAINE, MARYLAND, MASSACHUSETTS, MICHIGAN, MINNESOTA, MISSISSIPPI, MISSOURI, MONTANA, NEBRASKA, NEVADA, NEW HAMPSHIRE, NEW JERSEY, NEW MEXICO, NEW YORK, NORTH CAROLINA, NORTH DAKOTA, OHIO, OKLAHOMA, OREGON, PENNSYLVANIA, PUERTO RICO, RHODE ISLAND, SOUTH CAROLINA, SOUTH DAKOTA, TENNESSEE, TEXAS, UTAH, VERMONT, VIRGINIA, WASHINGTON, WEST VIRGINIA, WISCONSIN, WYOMING and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Georgia, Germany, Greece, Hungary, Ireland, Israel, Italy , Japan, Korea, Kuwait, Lithuania, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Reunion Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom.
Product codes and lots
05415067020215/ All Serial Numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2367-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2358-2024Class IIOngoingVoluntary: Firm initiated
Z-2359-2024Class IIOngoingVoluntary: Firm initiated
Z-2360-2024Class IIOngoingVoluntary: Firm initiated
Z-2361-2024Class IIOngoingVoluntary: Firm initiated
Z-2362-2024Class IIOngoingVoluntary: Firm initiated
Z-2363-2024Class IIOngoingVoluntary: Firm initiated
Z-2364-2024Class IIOngoingVoluntary: Firm initiated
Z-2365-2024Class IIOngoingVoluntary: Firm initiated
Z-2366-2024Class IIOngoingVoluntary: Firm initiated
Z-2367-2024Class IIOngoingVoluntary: Firm initiated
Z-2368-2024Class IIOngoingVoluntary: Firm initiated
Z-2369-2024Class IIOngoingVoluntary: Firm initiated