Endo Model M Tibial Components Modular Knee Prosthesis System. Product codes 15-2814/01, 15-2814/02...
Waldemar Link GmbH & Co. KG (Mfg Site) began a recall of Endo Model M Tibial Components Modular Knee Prosthesis System. Product codes 15-2814/01, 15-2814/02, 15-2814/03, 15-2814/04, 15-2815/11, 15-2815/12... on . FDA classified it as Class II and gives the reason as "Blind screws of the modular tibial component cannot be loosened intraoperatively,prolongation surgery due to intraoperatively change in procedure, probably to cementing technique". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2410-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Waldemar Link GmbH & Co. KG
- Product
- Endo Model M Tibial Components Modular Knee Prosthesis System. Product codes 15-2814/01, 15-2814/02, 15-2814/03, 15-2814/04, 15-2815/11, 15-2815/12, 15-2816/11, 15-2816/12, 15-2817/11, 15-2817/12, 15-2818/11, 15-2818/12, 15-2834/01, 15-2834/02, 15-2834/03, 15-2834/04, 15-2835/11, 15-2835/12, 15-2836/11, 15-2836/12, 15-2837/11, 15-2837/12, 15-2838/11, 15-2838/12, 15-3815/11, 15-3815/12, 15-3816/11, 15-3816/12, 15-3817/11, 15-3817/12, 15-3818/11, 15-3818/12, 15-8521/05, 15-8521/07, 15-8521/09, 15-8521/11, 15-8521/15, 15-8521/25, 15-8521/27, 15-8521/29, 15-8521/31, 15-8521/33
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- Blind screws of the modular tibial component cannot be loosened intraoperatively,prolongation surgery due to intraoperatively change in procedure, probably to cementing technique
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- LinkBio issue Urgent Medical Device Correction letter to Distributors via email on June 10, 2024. Letter states reason for recall, health risk and action to take: Please take note of the attached modified Surgical Technique for cementing the Tilastan Proximal Tibial Spacer or Proximal Tibial Segment, provided in Appendix A of this notice. No product return is required. Please respond using the attached Distributor Reply form to indicate your understanding of this Device Correction and of the need to apply the provided cementing surgical technique for the Tilastan Proximal Tibial Spacer or Proximal Tibial Segment, in case the blind screws are unable to be removed. Please contact LinkBio at 973-625-1333 or [email protected] for any questions
- Quantity in commerce
- 552 units; 7351 units OUS
- Distribution
- Nationwide Foreign: Algeria Argentina Australia Austria Belarus Belgium Brazil Bulgaria Canada China Colombia Czechia Denmark Ecuador Estonia Finland France Germany Greece Hungary India Indonesia Israel Italy Kenya Lithuania Luxembourg Mexico Netherlands Norway Peru Poland Romania Saudi Arabia Slovakia Slovenia Spain Sweden Switzerland Thailand T¿rkiye Ukraine United Arab Emirates United Kingdom Uruguay Vietnam
- Product codes and lots
- Product UDI-DI: 1 15-2814/01 04026575316243 2 15-2814/02 04026575316281 3 15-2814/03 04026575316298 4 15-2814/04 04026575316304 5 15-2815/11 04026575316007 6 15-2815/12 04026575316014 7 15-2816/11 04026575316021 8 15-2816/12 04026575316038 9 15-2817/11 04026575316052 10 15-2817/12 04026575316069 11 15-2818/11 04026575316076 12 15-2818/12 04026575316083 13 15-2834/01 04026575316731 14 15-2834/02 04026575316748 15 15-2834/03 04026575316755 16 15-2834/04 04026575316762 17 15-2835/11 04026575316557 18 15-2835/12 04026575316564 19 15-2836/11 04026575316571 20 15-2836/12 04026575316588 21 15-2837/11 04026575316595 22 15-2837/12 04026575316601 23 15-2838/11 04026575316618 24 15-2838/12 04026575316625 25 15-3815/11 04026575034727 26 15-3815/12 04026575034734 27 15-3816/11 04026575034741 28 15-3816/12 04026575034758 29 15-3817/11 04026575164042 30 15-3817/12 04026575164059 31 15-3818/11 04026575164028 32 15-3818/12 04026575164035 33 15-8521/05 04026575383207 34 15-8521/07 04026575383221 35 15-8521/09 04026575383238 36 15-8521/11 04026575383245 37 15-8521/13 04026575383269 38 15-8521/15 04026575383283 39 15-8521/25 04026575386406 40 15-8521/27 04026575386413 41 15-8521/29 04026575386420 42 15-8521/31 04026575386437 43 15-8521/33 04026575386444 44 15-8521/35 04026575386451 All Product lots manufactured since 01-Jun-2022.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2410-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2411-2024 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY