Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2411-2024Class II

Endo Model SL Tibial Components 16-2817/02 Tibial Component, Small, W: 60mm 16-2817/05 Tibial Component...

Waldemar Link GmbH & Co. KG / initiated 2024-05-22 / Open, Classified / root cause: manufacturing process

Waldemar Link GmbH & Co. KG (Mfg Site) began a recall of Endo Model SL Tibial Components 16-2817/02 Tibial Component, Small, W: 60mm 16-2817/05 Tibial Component, Medium, W: 65mm 16-2817/07 Tibial Component, Large, W... on . FDA classified it as Class II and gives the reason as "Blind screws of the modular tibial component cannot be loosened intraoperatively,prolongation surgery due to intraoperatively change in procedure, probably to cementing technique". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2411-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Waldemar Link GmbH & Co. KG
Product
Endo Model SL Tibial Components 16-2817/02 Tibial Component, Small, W: 60mm 16-2817/05 Tibial Component, Medium, W: 65mm 16-2817/07 Tibial Component, Large, W: 75mm 16-2817/32 Tibial Component Modular, Small, W: 60mm 16-2817/35 Tibial Component Modular, Medium, W: 65mm 16-2817/37 Tibial Component Modular, Large, W: 75mm
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Blind screws of the modular tibial component cannot be loosened intraoperatively,prolongation surgery due to intraoperatively change in procedure, probably to cementing technique
Root cause tag
manufacturing process
FDA root cause
Process design
Action
LinkBio issue Urgent Medical Device Correction letter to Distributors via email on June 10, 2024. Letter states reason for recall, health risk and action to take: Please take note of the attached modified Surgical Technique for cementing the Tilastan Proximal Tibial Spacer or Proximal Tibial Segment, provided in Appendix A of this notice. No product return is required. Please respond using the attached Distributor Reply form to indicate your understanding of this Device Correction and of the need to apply the provided cementing surgical technique for the Tilastan Proximal Tibial Spacer or Proximal Tibial Segment, in case the blind screws are unable to be removed. Please contact LinkBio at 973-625-1333 or [email protected] for any questions
Quantity in commerce
114 units US; 2351 units OUS
Distribution
Nationwide Foreign: Algeria Argentina Australia Austria Belarus Belgium Brazil Bulgaria Canada China Colombia Czechia Denmark Ecuador Estonia Finland France Germany Greece Hungary India Indonesia Israel Italy Kenya Lithuania Luxembourg Mexico Netherlands Norway Peru Poland Romania Saudi Arabia Slovakia Slovenia Spain Sweden Switzerland Thailand T¿rkiye Ukraine United Arab Emirates United Kingdom Uruguay Vietnam
Product codes and lots
Product UDI-DI 16-2817/02 04026575359202; 16-2817/05 04026575359219; 16-2817/07 04026575359226; 16-2817/32 04026575370870; 16-2817/35 04026575370887; 16-2817/37 04026575370894. All product lots manufactured since 01-Jun-2022.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2411-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2410-2024Class IIOngoingVoluntary: Firm initiated
Z-2411-2024Class IIOngoingVoluntary: Firm initiated