Endo Model SL Tibial Components 16-2817/02 Tibial Component, Small, W: 60mm 16-2817/05 Tibial Component...
Waldemar Link GmbH & Co. KG (Mfg Site) began a recall of Endo Model SL Tibial Components 16-2817/02 Tibial Component, Small, W: 60mm 16-2817/05 Tibial Component, Medium, W: 65mm 16-2817/07 Tibial Component, Large, W... on . FDA classified it as Class II and gives the reason as "Blind screws of the modular tibial component cannot be loosened intraoperatively,prolongation surgery due to intraoperatively change in procedure, probably to cementing technique". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2411-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Waldemar Link GmbH & Co. KG
- Product
- Endo Model SL Tibial Components 16-2817/02 Tibial Component, Small, W: 60mm 16-2817/05 Tibial Component, Medium, W: 65mm 16-2817/07 Tibial Component, Large, W: 75mm 16-2817/32 Tibial Component Modular, Small, W: 60mm 16-2817/35 Tibial Component Modular, Medium, W: 65mm 16-2817/37 Tibial Component Modular, Large, W: 75mm
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- Blind screws of the modular tibial component cannot be loosened intraoperatively,prolongation surgery due to intraoperatively change in procedure, probably to cementing technique
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- LinkBio issue Urgent Medical Device Correction letter to Distributors via email on June 10, 2024. Letter states reason for recall, health risk and action to take: Please take note of the attached modified Surgical Technique for cementing the Tilastan Proximal Tibial Spacer or Proximal Tibial Segment, provided in Appendix A of this notice. No product return is required. Please respond using the attached Distributor Reply form to indicate your understanding of this Device Correction and of the need to apply the provided cementing surgical technique for the Tilastan Proximal Tibial Spacer or Proximal Tibial Segment, in case the blind screws are unable to be removed. Please contact LinkBio at 973-625-1333 or [email protected] for any questions
- Quantity in commerce
- 114 units US; 2351 units OUS
- Distribution
- Nationwide Foreign: Algeria Argentina Australia Austria Belarus Belgium Brazil Bulgaria Canada China Colombia Czechia Denmark Ecuador Estonia Finland France Germany Greece Hungary India Indonesia Israel Italy Kenya Lithuania Luxembourg Mexico Netherlands Norway Peru Poland Romania Saudi Arabia Slovakia Slovenia Spain Sweden Switzerland Thailand T¿rkiye Ukraine United Arab Emirates United Kingdom Uruguay Vietnam
- Product codes and lots
- Product UDI-DI 16-2817/02 04026575359202; 16-2817/05 04026575359219; 16-2817/07 04026575359226; 16-2817/32 04026575370870; 16-2817/35 04026575370887; 16-2817/37 04026575370894. All product lots manufactured since 01-Jun-2022.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2410-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2411-2024 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.