DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes...
Beckman Coulter Mishima K.K. began a recall of DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate... on . FDA classified it as Class II and gives the reason as "There are two software bug issues with the analyzer. Issue 1: When a dedicated rack is reserved for calibration, after a calibration test has been placed and completed, the same rack may be used to place both a patient...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2455-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Beckman Coulter Mishima K.K.
- Product
- DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- There are two software bug issues with the analyzer. Issue 1: When a dedicated rack is reserved for calibration, after a calibration test has been placed and completed, the same rack may be used to place both a patient order and a calibration order. When a patient sample and a non-barcoded calibrator tube is processed on the rack, the calibration result will be reported as a patient result. Issue 2: calibration with expired calibrator. They both may cause erroneous patient results, but the probability of serious adverse health consequences or death is unlikely.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- During the week of June 20, 2024, the firm sent an "URGENT MEDICAL DEVICE RECALL" Letter via email/postal mail to customers informing them of two software related issues associated with use of the Clinical Chemistry Analyzer. Issue 1: When a dedicated rack is reserved for calibration, after a calibration test has been placed and completed, the same rack may be used to place both a patient order and a calibration order. When a patient sample and a non-barcoded calibrator tube is processed on the rack, the calibration result will be reported as a patient result. This may cause a false low result or a false high result to be reported to a physician. Issues 2: When a calibration order is manually placed (calibrator lot is not expired) and then the calibrator expires before the calibrator material is placed onto the analyzer for processing, the analyzer will process the expired calibration lot and generate patient results that may be inaccurate, potentially causing a false low result or a false high result to be reported to a physician. Customer are instructed to: For Issue 1 -Follow the IFU Note in Assigning Racks or Calibrator IDs to Calibrators section on pg. 8-7. -Alternatively, change calibrator Load Type to Bar Code Configuration and use barcoded containers for calibration. -Refer to IFU Assigning Racks or Calibrator IDs to Calibrators section on pg. 8-6, and in the software user interface Calibration section, under Calibrations Due Now, review the Load Type configuration, rack ID and cup position set up for the calibrators. Issue 2: -Verify that the calibrator has not expired prior to loading the calibrator material onto the analyzer for processing. -Cancel calibration orders after the calibrator lot material expires. -Repeat any test results generated from a calibration curve that used expired calibrator lot material. For questions/assistance-contact Customer Support Center at http://www.beckmancoulter.com
- Quantity in commerce
- 24 analyzers
- Distribution
- Worldwide - U.S. Nationwide distribution in the states of AL, CA, GA, HI, IL, MI, MN, MS, NE, NY, OR, PA and SC. The countries of Lebanon, and South Africa.
- Product codes and lots
- Catalog Number: C63520 UDI-DI code: 14987666545065 Serial Numbers: 2023070010 2023060005 2023090015 2023070009 2023080011 2023080012 2023100019 2023100018 2023100020 2024010037 2024010036 2023110022 2024010038 2023090016 2024020040 2024020041 2024020039 2024020043 2024020042 2023110023 2024020045 2024020046 2024010024 2024010025
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2455-2024 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
