Z-2319-2024Class I

Impella CP with SmartAssist, Product No.: 1000080 (REF 0048-0003). left heart support blood pump

Abiomed, Inc. / initiated 2024-05-31 / Open, Classified

Abiomed, Inc. began a recall of Impella CP with SmartAssist, Product No.: 1000080 (REF 0048-0003). left heart support blood pump on . FDA classified it as Class I and gives the reason as "Nine (9) Impella CP pumps failed inspection and were inadvertently released". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2319-2024
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Abiomed, Inc.
Product
Impella CP with SmartAssist, Product No.: 1000080 (REF 0048-0003). left heart support blood pump
Product code
OZD Temporary Non-Roller Type Left Heart Support Blood Pump
Reason for recall
Nine (9) Impella CP pumps failed inspection and were inadvertently released.
FDA root cause
Release of Material/Component prior to receiving test results
Action
On May 31, 2024 consignees were contacted by phone with a script followed-up by sending URGENT VOLUNTARY MEDICAL DEVICE RECALL (REMOVAL) letters to consignees. ACTIONS TO BE TAKEN BY THE CUSTOMER/USER: 1.Examine your inventory immediately to determine if you have product subject to this recall(removal). QUARANTINE AND DO NOT USE THE SUBJECT PRODUCTS. 2.Your Abiomed Representatives have been provided instructions to support facilitation of locating, separating, and returning affected product for replacement. Work with your Abiomed Representative toreview, complete all fields, and sign the Business Response Form attached on page 4. IMPORTANT:The Business Response Form must be completed even if you have used any units of the subjectproduct on hand. 3.Forward this notice to any personnel in your facility who need to be informed. 4.If any of the product subject to this recall (removal) has been forwarded to another facility, contact thatfacility and provide a copy of this notice to the relevant personnel. 5.Post a copy of this notice in a visible area for awareness of this field safety notice. 6.As with any medical device, adverse reactions or quality problems experienced with the use of this productshould be reported to the FDA s MedWatch Adverse Event Reporting Program using the link below: https://www.fda.gov/safety/medwatch-fda-safety-information-and-adverse-event-reporting- program/reporting-serious-problems-fda Please contact Abiomed Customer Service, 1-800-422-8666, option 2 (email [email protected] or contact your local clinical field staff if you have questions or concerns regarding this notice.
Quantity in commerce
9 units
Distribution
Domestic only: FL, MA, OH TX.
Product codes and lots
Product No.: 1000080 (REF 0048-0003); UDI/DI: 00813502012279; Lot/Serial No.: Lot 1798046; Serial numbers: 504354, 504355, 504356, 504357, 504359, 504360, 504361, 504362, 504363.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2319-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2319-2024Class IOngoingVoluntary: Firm initiated