Z-2412-2024Class II

Surgical Image Guidance Functionality, software used with NGENUITY 3D Visualization System, Part Number...

Alcon Laboratories / initiated 2024-05-23 / Open, Classified / root cause: software

Alcon Research LLC began a recall of Surgical Image Guidance Functionality, software used with NGENUITY 3D Visualization System, Part Number: 8065000399 on . FDA classified it as Class II and gives the reason as "After surgery initiation, while using surgical image guidance, if significant eye rotation/movement occurs, a software anomaly may cause re-establishment of incorrect new registration angle, range indicator will display...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-2412-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Alcon Laboratories
Product
Surgical Image Guidance Functionality, software used with NGENUITY 3D Visualization System, Part Number: 8065000399
Product code
FWJ Camera, Cine, Microsurgical, Without Audio
Reason for recall
After surgery initiation, while using surgical image guidance, if significant eye rotation/movement occurs, a software anomaly may cause re-establishment of incorrect new registration angle, range indicator will display incorrectly, which could result in incorrect placement of toric IOL axis, which could cause astigmatic error under/over correction resulting in decreased uncorrected visual acuity.
Root cause tag
software
FDA root cause
Software design
Action
On 5/23/24, correction notices were mailed to customers informing them of the following: Prior to ending procedure, re-perform registration to ensure proper axis alignment. Firm is expediting a software update that is expected to be released to upgrade systems running on software version 1.5 starting mid-June 2024. Firm's Technical Service representative will contact you to schedule your service appointment for this software upgrade. If you choose to use the image guidance functionality for toric alignment, to mitigate the potential for this anomaly, users are advised to re-perform registration prior to ending the case or if any axis discrepancy is noted during the case. To re-perform registration, go to the registration step, re-confirm the reference image to the new surgery image, adjust the registration angle (if necessary), select "Confirm" and proceed to toric alignment or desired step. Complete and return the response form via email to [email protected] If you have any questions regarding this issue, please contact your Firm representative or Technical Support at 1-800-832-7827 or contact your local Sales Representative.
Quantity in commerce
230
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, CA, CO, CT, FL, GA, HI, IA, IL, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NJ, NV, NY, OH, OK, PA, SC, TN, TX, WA, WI, PR and the countries of Canada, Japan, France.
Product codes and lots
UDI-DI: 00380650003997. Software Version: 1.5. Serial Numbers: 004-0680001, 004-0680002, 004-0680003, 004-0680004, 004-0680007, 004-0680008, 004-0680017, 004-0680026, 004-0680036, 004-0680047, 004-0680048, 004-0680086, 004-0680090, 004-0680092, 004-0680096, 004-0680100, 004-0680107, 004-0680112, 004-0680120 004-0680127,004-0680131, 004-0680134, 004-0680140, 004-0680152, 004-0680153, 004-0680161, 004-0680163, 004-0680188, 004-0680190,004-0680203, 004-0680206, 004-0680266, 004-0680267, 004-0680271, 004-0680273, 004-0680274, 004-0680283, 004-0680288, 004-0680293, 004-0680294, 004-0680306, 004-0680311, 004-0680315, 004-0680319, 004-0680325, 004-0680334, 004-0680336, 004-0680337, 004-0680338, 004-0680339, 004-0680346, 004-0680348, 004-0680352, 004-0680362, 004-0680404, 004-0680417, 004-0680418, 005-7230003, 005-7230005, 005-7230007, 005-7230017, 005-7230019, 005-7230021, 005-7230022, 005-7230023, 005-7230031, 005-7230035, 005-7230057, 005-7230081, 005-7230085, 005-7230086, 005-7230102, 005-7230108, 005-7230109, 005-7230186, 007-4610018, 007-4610023, 007-4610030, 007-4610032, 007-4610033, 007-4610034, 007-4610037, 007-4610044, 007-4610074, 007-4610107, 007-4610112, 007-4610115, 007-4610122, 007-4610123, 007-4610124, 007-4610125, 007-4610127, 007-4610132, 007-4610138, 007-4610166, 007-4610167, 007-4610179, 007-4610181, 007-4610186, 007-4610217, 007-4610219, 007-4610220, 007-4610226, 007-4610229, 007-4610230, 007-4610262, 007-4610275, 007-4610283, 007-4610287, 007-4610288, 007-4610308, 007-4610309, 007-4610351, 007-4610355, 007-4610359, 007-4610360, 007-4610366, 007-4610394, 007-4610404, 007-4610405, 007-4610406, 007-4610409, 009-4570001, 009-4570004, 009-4570005, 009-4570007, 009-4570013, 009-4570014, 009-4570015, 009-4570016, 009-4570019, 009-4570022, 009-4570033, 009-4570035, 009-4570044, 009-4570047, 009-4570048, 009-4570049, 009-4570051,009-4570052, 009-4570053 009-4570054,009-4570057, 009-4570059, 009-4570060, 009-4570064, 009-4570068, 009-4570070, 009-4570072, 009-4570073, 009-4570096, 009-4570097, 009-4570099, 009-4570100, 009-4570114, 009-4570116, 009-4570125, 009-4570127, 009-4570129, 009-4570130, 009-4570136, 009-4570137, 009-4570144, 009-4570148, 009-4570149, 009-4570150, 009-4570153, 009-4570155, 009-4570156, 009-4570158, 009-4570160, 009-4570161, 009-4570167, 009-4570171, 009-4570173, 009-4570176,009-4570178, 009-4570179, 009-4570180, 009-4570183, 009-4570184, 009-4570200, 009-4570204, 009-4570207, 009-4570233, 009-4570249, 009-4570256, 009-4570259, 009-4570263, 009-4570270, 009-4570275, 009-4570278, 009-4570281, 009-4570289, 009-4570293, 009-4570294, 009-4570300, 009-4570308, 009-4570317, 009-4570320, 009-4570331, 009-4570342, 009-4570343, 009-4570350, 009-4570353, 009-4570357, 009-4570365, 009-4570368, 009-4570393, 009-4570397, 009-4570425, 009-4570431, 009-4570435, 009-4570440, 009-4570448, 009-4570449, 009-4570451, 009-4570455, 009-4570457, 009-4570459, 009-4570460, 009-4570474, 009-4570489, 009-4570508, 009-4570510, DM550021, DM550024, DM550043, DM550051, DM550054

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2412-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2412-2024Class IIOngoingVoluntary: Firm initiated