Z-2544-2024Class II

Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Catalog #C11137

Siemens Healthcare Diagnostics Inc. / initiated 2024-05-22 / Open, Classified / root cause: software

Beckman Coulter, Inc. began a recall of Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Catalog #C11137 on . FDA classified it as Class II and gives the reason as "The reason for this recall is Beckman Coulter has determined that when a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS); is running with the software versions 1.17 and below...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-2544-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
The reason for this recall is Beckman Coulter has determined that when a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS); is running with the software versions 1.17 and below; and a sample tube is sent from the LAS to DxI 9000 Access Immunoassay Analyzer, an out of sequence message will be sent from the Analyzer to the LAS. This issue may result in multiple error messages (i.e., a sample pick and place (PnP) error or a no result obtained (NRT) flag may be displayed) and subsequently a risk of delays in processing/reporting patient test results and patient treatment.
Root cause tag
software
FDA root cause
Software design
Action
The recalling firm issued letters dated 5/16/2024 on 5/22/2024 via mail and email. The letter provided the issue, impact, and actions to take. The actions included: (1) If the issue occurs, initialize the system as prompted by the analyzer; (2) Contact the Beckman Coulter representative and schedule a service visit if the laboratory has experienced sample PnP or aliquot pipettor error that is not resolved by initializing the system; (3) If samples received an NRT flag, request the test again and reload the sample; (4) Identify any sample tubes that were not processed and reload them on the Dxl 9000 Access Immunoassay Analyzer from the sample handler input area. The letter informs the consignee that Beckman Coulter has identified the root cause of the issue and will implement the correction with a future software release. Beckman Coulter will contact the consignee to schedule the software upgrade when it is available.
Quantity in commerce
183 devices
Distribution
Worldwide distribution - US Nationwide distribution in the states of CA, FL, GA, IN, KS, ME, NH, NJ, OK, PA, and TX. The countries of Australia, Austria, Bahrain, Belgium, Brazil, Croatia, Czech Republic, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Switzerland, Taiwan, and United Kingdom.
Product codes and lots
Software versions 1.17 and below, UDI-DI 15099590732103.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2544-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2544-2024Class IIOngoingVoluntary: Firm initiated