Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Catalog #C11137
Beckman Coulter, Inc. began a recall of Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Catalog #C11137 on . FDA classified it as Class II and gives the reason as "The reason for this recall is Beckman Coulter has determined that when a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS); is running with the software versions 1.17 and below...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2544-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- The reason for this recall is Beckman Coulter has determined that when a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS); is running with the software versions 1.17 and below; and a sample tube is sent from the LAS to DxI 9000 Access Immunoassay Analyzer, an out of sequence message will be sent from the Analyzer to the LAS. This issue may result in multiple error messages (i.e., a sample pick and place (PnP) error or a no result obtained (NRT) flag may be displayed) and subsequently a risk of delays in processing/reporting patient test results and patient treatment.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The recalling firm issued letters dated 5/16/2024 on 5/22/2024 via mail and email. The letter provided the issue, impact, and actions to take. The actions included: (1) If the issue occurs, initialize the system as prompted by the analyzer; (2) Contact the Beckman Coulter representative and schedule a service visit if the laboratory has experienced sample PnP or aliquot pipettor error that is not resolved by initializing the system; (3) If samples received an NRT flag, request the test again and reload the sample; (4) Identify any sample tubes that were not processed and reload them on the Dxl 9000 Access Immunoassay Analyzer from the sample handler input area. The letter informs the consignee that Beckman Coulter has identified the root cause of the issue and will implement the correction with a future software release. Beckman Coulter will contact the consignee to schedule the software upgrade when it is available.
- Quantity in commerce
- 183 devices
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CA, FL, GA, IN, KS, ME, NH, NJ, OK, PA, and TX. The countries of Australia, Austria, Bahrain, Belgium, Brazil, Croatia, Czech Republic, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Switzerland, Taiwan, and United Kingdom.
- Product codes and lots
- Software versions 1.17 and below, UDI-DI 15099590732103.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2544-2024 | Class II | Ongoing | Voluntary: Firm initiated |