Z-2415-2024Class II

ESG-410 Electrosurgical Generator- In conjunction with electrosurgical accessories intended for cutting and...

OLYMPUS Corporation of the Americas / initiated 2024-06-21 / Open, Classified / root cause: design

Olympus Corporation of the Americas began a recall of ESG-410 Electrosurgical Generator- In conjunction with electrosurgical accessories intended for cutting and coagulation of tissue in open, laparoscopic , and... on . FDA classified it as Class II and gives the reason as "Electrosurgical Generators ESG-410 does not power on upon use, may result in delays in initiating treatment or prolonged surgery". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-2415-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
OLYMPUS Corporation of the Americas
Product
ESG-410 Electrosurgical Generator- In conjunction with electrosurgical accessories intended for cutting and coagulation of tissue in open, laparoscopic , and endoscopic surgeries Model/Catalog Number: WA91307C
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Electrosurgical Generators ESG-410 does not power on upon use, may result in delays in initiating treatment or prolonged surgery
Root cause tag
design
FDA root cause
Component design/selection
Action
Olympus issued URGENT: MEDICAL DEVICE REMOVAL Letter on 6/21/24. Letter states reason for recall, health risk and action to take: An Olympus representative will reach out to you to coordinate a repair to replace the power supply on your affected ESG-410 unit(s). Technical analysis of this issue performed by Olympus has determined that when the device is successfully powered on, it will work within specification. Therefore, you may continue to use your ESG-410 generator until the power supply is replaced. To minimize procedural delays, power on the generator prior to starting the procedure and, whenever possible, have a replacement device available. Additionally, Olympus requires you to take the following actions: 1. Carefully read the content of this notification. 2. If you have further distributed this product, identify your customers, and forward them this notification. 3. Acknowledge receipt of this letter through the Olympus web portal: a. Go to https://olympusamerica.com/recall Olympus requests that you report any complaints, including the ESG-410 generator not powering on, to the Technical Assistance Center (TAC) at 1- 800-848-9024 (option 1). If you require additional information, please do not hesitate to contact me by phone at (647) 999-3203 from Monday through Friday, 9 am to 5 pm EDT, or by e-mail at [email protected].
Quantity in commerce
86
Distribution
US Nationwide distribution.
Product codes and lots
UDI-DI: 04042761086349 Serial Numbers: 102410, 102409, 102408, 102419, 102418, 102417, 102416, 102420, 102412, 102407, 102394, 102393, 102401, 102400, 102405, 102404, 102403, 102428, 102424, 102422, 102438, 102432, 102431, 102430, 102429, 102437, 102435, 102433, 102442, 102440, 102439, 102436, 102366, 102364, 102372, 102371, 102444, 102441, 102502, 102499, 102498, 102497, 102515, 102513, 102512, 102511, 102506, 102504, 102510, 102505, 102503, 102501, 102500, 102522, 102520, 102518, 102517, 102516, 102523, 102519, 102396, 102369, 102434, 102392, 102823, 102822, 102821, 102819, 102818, 102802, 102814, 102810, 102809, 102808, 102834, 102830, 102829, 102828, 102919, 102925, 102917, 102916, 102913, 102912, 102836, 102921

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2415-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2415-2024Class IIOngoingVoluntary: Firm initiated