ESG-410 Electrosurgical Generator- In conjunction with electrosurgical accessories intended for cutting and...
Olympus Corporation of the Americas began a recall of ESG-410 Electrosurgical Generator- In conjunction with electrosurgical accessories intended for cutting and coagulation of tissue in open, laparoscopic , and... on . FDA classified it as Class II and gives the reason as "Electrosurgical Generators ESG-410 does not power on upon use, may result in delays in initiating treatment or prolonged surgery". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2415-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- ESG-410 Electrosurgical Generator- In conjunction with electrosurgical accessories intended for cutting and coagulation of tissue in open, laparoscopic , and endoscopic surgeries Model/Catalog Number: WA91307C
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- Electrosurgical Generators ESG-410 does not power on upon use, may result in delays in initiating treatment or prolonged surgery
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Olympus issued URGENT: MEDICAL DEVICE REMOVAL Letter on 6/21/24. Letter states reason for recall, health risk and action to take: An Olympus representative will reach out to you to coordinate a repair to replace the power supply on your affected ESG-410 unit(s). Technical analysis of this issue performed by Olympus has determined that when the device is successfully powered on, it will work within specification. Therefore, you may continue to use your ESG-410 generator until the power supply is replaced. To minimize procedural delays, power on the generator prior to starting the procedure and, whenever possible, have a replacement device available. Additionally, Olympus requires you to take the following actions: 1. Carefully read the content of this notification. 2. If you have further distributed this product, identify your customers, and forward them this notification. 3. Acknowledge receipt of this letter through the Olympus web portal: a. Go to https://olympusamerica.com/recall Olympus requests that you report any complaints, including the ESG-410 generator not powering on, to the Technical Assistance Center (TAC) at 1- 800-848-9024 (option 1). If you require additional information, please do not hesitate to contact me by phone at (647) 999-3203 from Monday through Friday, 9 am to 5 pm EDT, or by e-mail at [email protected].
- Quantity in commerce
- 86
- Distribution
- US Nationwide distribution.
- Product codes and lots
- UDI-DI: 04042761086349 Serial Numbers: 102410, 102409, 102408, 102419, 102418, 102417, 102416, 102420, 102412, 102407, 102394, 102393, 102401, 102400, 102405, 102404, 102403, 102428, 102424, 102422, 102438, 102432, 102431, 102430, 102429, 102437, 102435, 102433, 102442, 102440, 102439, 102436, 102366, 102364, 102372, 102371, 102444, 102441, 102502, 102499, 102498, 102497, 102515, 102513, 102512, 102511, 102506, 102504, 102510, 102505, 102503, 102501, 102500, 102522, 102520, 102518, 102517, 102516, 102523, 102519, 102396, 102369, 102434, 102392, 102823, 102822, 102821, 102819, 102818, 102802, 102814, 102810, 102809, 102808, 102834, 102830, 102829, 102828, 102919, 102925, 102917, 102916, 102913, 102912, 102836, 102921
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2415-2024 | Class II | Ongoing | Voluntary: Firm initiated |