BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, for FILMARRAY Systems, when used in...
BioFire Diagnostics, LLC began a recall of BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, for FILMARRAY Systems, when used in conjunction with Culture Media Bottles... on . FDA classified it as Class II and gives the reason as "If blood culture identification panel is used in conjunction with specific lots of culture media bottles, then false positive Serratia marcescens results may occur, due to an increased level of non-viable organism from...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2456-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- BioFire Diagnostics, LLC
- Product
- BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, for FILMARRAY Systems, when used in conjunction with Culture Media Bottles: BACT/ALERT FA PLUS, REF: 410851, and BACT/ALERT PF PLUS, REF: 410853
- Product code
- PEN Gram-Negative Bacteria And Associated Resistance Markers
- Reason for recall
- If blood culture identification panel is used in conjunction with specific lots of culture media bottles, then false positive Serratia marcescens results may occur, due to an increased level of non-viable organism from serratia marcescens targets in culture media bottles, false positive result may lead to an inappropriate change in patient therapy.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 06/19/24, recall notices were distributed to laboratory managers who were informed the following: 1) Positive results for Serratia marcescens should be confirmed by another method prior to reporting the test results. 2) Distribute this information to all appropriate personnel in your laboratory, retain a copy in your files, post this letter in or near the laboratory, and forward this information to all parties that may use this product, including others to whom you may have transferred our product. 3) Complete and return the acknowledgement form via email to [email protected] If you require additional assistance or have any questions, please contact firm's US Commercial Operations. at (800) 682-2666. For non-urgent inquiries, please contact the Technical Support Center by email at [email protected]. The anticipated turnaround time for email inquiries is 24 - 48 hours during normal business hours. On 11/04/24, additional notices were sent to customers informing them of three additional affected lots. On 2/3/2025, additional notices were sent to customers informing them of one additional affected lot. On 6/3/2025, additional notices were sent to customers informing them of five additional affected lots.
- Quantity in commerce
- 99,822
- Distribution
- Worldwide - US Nationwide distribution including in the states of FL, IL, KY, CO, KS, NC, AK, NY, ND, CA, TX, MD, GA, OH, WI, SD, LA, MT, VA, IA, MO, MA, WV, NE, OK, AZ, OR, SC, MI, IN, ID, TN, WA, NM, NJ, AR, AL, MN, NH, NV, PA, WY, MS, CT and the countries of AE, AM, AR, AT, BA, BG, BH, BN, BR, CA, CH, CI, CK, CN, CO, CZ, DE, DZ, EC, EE, EG, ES, FI, FR, GA, GB, GE, GR, GT, HR, HU, IL, IN, IQ, IS, IT, JO, JP, KK, KR, KW, LB, LT, LV, MK, MM, MX, MY, NG, NO, NW, OM, PA, PH, PK, PL, PT, QA, RO, RS, SA, SI, TN, TR, TW, UG, UY, VN, WF.
- Product codes and lots
- BCID2 Panel REF: RFIT-ASY-0147, UDI-DI: 00815381020338, All unexpired lots. BACT/ALERT REF/UDI-DI: Lot/Expiration: 410851/03573026596057: 0004101638/3/20/2024, 0004101930/7/28/2024, 0004102026/9/4/2024, 0004102408/2/16/2025, 0004102996/9/24/2025, 0004102946/8/29/2025, 0004102956/9/17/2025, 0004102998/9/24/2025. 410853/03573026596095: 0004101718/4/24/2024, 0004101790/6/1/2024, 0004101958/8/3/2024, 0004102452/3/3/2025, 0004102964/9/19/2025
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2456-2024 | Class II | Ongoing | Voluntary: Firm initiated |