Brand Name: Perfusor Space Product Name: Perfusor Space Infusion Pump System Model/Catalog Number: 8713030U...
B Braun Medical Inc began a recall of Brand Name: Perfusor Space Product Name: Perfusor Space Infusion Pump System Model/Catalog Number: 8713030U Product Description: Perfusor Space Infusion Pump... on . FDA classified it as Class II and gives the reason as "Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2433-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- B Braun Medical Inc
- Product
- Brand Name: Perfusor Space Product Name: Perfusor Space Infusion Pump System Model/Catalog Number: 8713030U Product Description: Perfusor Space Infusion Pump System, Infusion Pump, 1 each Component: N/A
- Product code
- FRN Pump, Infusion
- Reason for recall
- Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On June 14, 2024, the firm notified customers via URGENT MEDICAL DEVICE FIELD CORRECTION NOTIFICATION letters. Customers were advised to not use the recalled syringes in Perfusor Space Syringe Pumps. All users should be informed of the correction. Customers were provided with an attachment to post in care areas as a reminder to not use the recalled syringes with the specified pumps.
- Quantity in commerce
- 14949
- Distribution
- Domestic distribution nationwide. International distribution to Canada.
- Product codes and lots
- Lot Code: Model No 8713030U; UDI-DI 4046963716745; All Serial Numbers
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2433-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2434-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2435-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2436-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2437-2024 | Class II | Ongoing | Voluntary: Firm initiated |