Z-2545-2024Class II

Discrete photometric chemistry analyzer for clinical use

Siemens Healthcare Diagnostics Inc. / initiated 2024-06-14 / Open, Classified

Beckman Coulter, Inc. began a recall of Discrete photometric chemistry analyzer for clinical use on . FDA classified it as Class II and gives the reason as "Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2545-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Discrete photometric chemistry analyzer for clinical use.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module. Customers affected by this issue may observe an increase in reagent carousel motion errors or QSMC errors that occur while unloading or loading reagent packs during system initialization and sample processing, which causes a non-operational system status and cancels tests; and patient results may be delayed if these errors cancel test processing.
FDA root cause
Under Investigation by firm
Action
The recall notification was sent to consignees during the week of 06/14/2024 via email and postal mail. Consignees were instructed to follow the troubleshooting instructions in section 6.4 of the Access 2 Reference Manual and contact the Beckman Coulter representative if the laboratory has experienced repeated reagent carousel motion or QSMC errors. Consignees are asked to respond within 10 days indicating they have received the important communication either via by email or by returning the response form.
Quantity in commerce
59 units
Distribution
Worldwide - US Nationwide distribution in the states of NE, MT, LA, TX, NY, CA, MI, AL, IL the countries of Austria, Belgium, Georgia, India, Indonesia, Italy, Korea, Malaysia, Myanmar, Nigeria, Switzerland, Vietnam.
Product codes and lots
Catalog Number: 81600N, C11907; UDI: 150995901; Serial Numbers: 574415, 574440, 574484,574416, 574441, 574485, 574417, 574442, 574486, 574418, 574443, 574488,574419, 574444, 574489,574420, 574445, 574490, 574421, 574446, 574491, 574422, 574447 574494,574423, 574448, 574497,574424, 574454, 574502, 574425, 574455, 574503, 574426, 574456 574504,574427, 574457, 574505,574428, 574459, 574506, 574429, 574469, 574507, 574430, 574470 574509,574432, 574471, 574513, 574433, 574472, 574514, 574434, 574477, 574515, 574435, 574479

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2545-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2545-2024Class IIOngoingVoluntary: Firm initiated