Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2454-2024Class II

SOFT-VU KUMPE 5F X 65CM -Used for patients in need of angiographic diagnosis. Catalog Number: 10732701

Boston Scientific Corporation / initiated 2024-06-18 / Open, Classified / root cause: packaging

Angiodynamics, Inc. began a recall of SOFT-VU KUMPE 5F X 65CM -Used for patients in need of angiographic diagnosis. Catalog Number: 10732701 on . FDA classified it as Class II and gives the reason as "Mislabeled. The shelf-box labeling (unit of sale) is accurate for the packaged devices, however the inner pouch label may contain information for a different model (catheter length)". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2454-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Boston Scientific Corporation
Product
SOFT-VU KUMPE 5F X 65CM -Used for patients in need of angiographic diagnosis. Catalog Number: 10732701
Product code
DQO Catheter, Intravascular, Diagnostic
Reason for recall
Mislabeled. The shelf-box labeling (unit of sale) is accurate for the packaged devices, however the inner pouch label may contain information for a different model (catheter length)
Root cause tag
packaging
FDA root cause
Packaging change control
Action
Angiodynmaics issued Urgent Medical Device Recall Letter on 6/18/24 via Federal Express to end-users. Letter states reason for recall, health risk and action to take: " IMMEDIATELY o Stop using the product subject to recall. o Remove any affected (recalled) product from your inventory (whether in labs, Central Supply, Shipping and Receiving or ANY other location). o Segregate this product in a secure location for return to AngioDynamics, Inc. o Forward a copy of this recall notification to all sites to which you have distributed affected product. 2. Complete and return the Reply Verification Tracking Form. Promptly complete, sign, and return the enclosed Reply Verification Tracking Form (even if you do not have any product to return); following the directions on this page and the Reply Verification Tracking Form. Email Reply Verification Tracking Form (preferred): [email protected] Fax Reply Verification Tracking Form: Attn: Soft-Vu Kumpe Recall Coordinator Fax number 1-855-273-0519 " Call AngioDynamics Customer Service at 1-800-772-6446 between 8:00 a.m. and 6:00 p.m. (Monday Friday: Eastern Standard Time) to obtain a replacement or credit for your returned product. " Report any Adverse reactions or quality problems associated with recalled devices to US FDA via MedWatch Adverse Event Reporting program either online, by regular mail or by fax. 3. Package and Return the Recalled Product. " Package any product that is being returned in an appropriate shipping box. " Write the RMA number on the RMA/Address label (provided on the Recall Verification Tracking Form) and affix the label to the outside of the shipping box. " Seal the box and return to: AngioDynamics, Inc. 24 Native Drive Queensbury, NY 12804 Attn: Soft-Vu Kumpe Recall Coordinator
Quantity in commerce
595 boxes
Distribution
US Nationwide distribution.
Product codes and lots
UPN: H787107327015 UDI-DI: 25051684009772 (Box) 15051684009775 (Pouch) Lot Number: A0424031

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2454-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2453-2024Class IIOngoingVoluntary: Firm initiated
Z-2454-2024Class IIOngoingVoluntary: Firm initiated