Multiple Digital Analog Products labeled as: NobelBiocare Tri-Lobe-compatible 6.0mm Digital Analog, REF...
Preat Corp began a recall of Multiple Digital Analog Products labeled as: NobelBiocare Tri-Lobe-compatible 6.0mm Digital Analog, REF 9000703 Straumann Tissue-Level-compatible WN Digital... on . FDA classified it as Class II and gives the reason as "Due to a manufacturing issue, the product has a out of specification rotation of the digital analog used to visually inspect dental abutments and restorations". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2687-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- PREAT Corporation
- Product
- Multiple Digital Analog Products labeled as: NobelBiocare Tri-Lobe-compatible 6.0mm Digital Analog, REF 9000703 Straumann Tissue-Level-compatible WN Digital Analog, REF 9001303 3i Certain-compatible 5.0mm Digital Analog, REF 9001903 3i Certain-compatible 6.0mm Digital Analog, REF 9002003 Astra-compatible 3.0mm Digital Analog, REF 9002103 Astra-compatible Aqua 3.5/4.0mm Digital Analog, REF 9002203 Astra-compatible Lilac 4.5/5.0mm Digital Analog, REF 9002203 BioHorizons-compatible 3.0mm Digital Analog, REF 9005203 Legacy-compatible 3.0mm Digital Analog, REF 9005303 Digital Analog - BioHorizons - 3.5mm, Digital Analog - BioHorizons - 4.5mnm, REF 9005403 Legacy-compatible 3.5mm Digital Analog, REF 9005503 Legacy-compatible 5.7mm Digital Analog, REF 9008503 Astra EV 4.8mm Milled Titanium Abutment REF 9006767 Digital analogs are used for visual inspection of abutments and restoration in the dental laboratory.
- Product code
- NDP Accessories, Implant, Dental, Endosseous
- Reason for recall
- Due to a manufacturing issue, the product has a out of specification rotation of the digital analog used to visually inspect dental abutments and restorations.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On 07/22/2024, the firm sent an "URGENT: MEDICAL DEVICE RECALL UPDATE" Letter via certified mail informing customers that Digital Analogs were manufactured out of specification for the rotational feature which may result in the dental restoration/implant may be slightly different than the position of the actual implant in the patient's mouth. Customers are instructed to: Product in the scope of this recall should be discarded or returned to PREAT CORPORATION. For questions contact Quality Regulatory Affairs at 224-622-7191 or email [email protected]
- Quantity in commerce
- 185 devices
- Distribution
- U.S. Nationwide distribution in the states of AK, AZ, FL, GA, IL, LA, MD, MI, MN, NY, OH, OK, OR, PA, RI, TX, UT, and WA
- Product codes and lots
- Model Numbers: 9000703, 9001303, 9001903, 9002003, 9002103, 9002203, 9002203, 9005203, 9005203, 9005303, 9005403, 9005503, 9008503, 9006767 UDI-DI codes: 00842092178690, 00842092181898, 00842092181959, 00842092178737, 00842092178775, 00842092178812, 00842092178812, 00842092178898, 00842092178898, 00842092178997, 00842092178928, 00842092178959, 00842092182130 Lot Numbers: 250219, 248890, 251432, 250602, 250418, 250213, 251459, 249071, 250335, 250223, 250336, 249510, 250215.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2687-2024 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.