Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2687-2024Class II

Multiple Digital Analog Products labeled as: NobelBiocare Tri-Lobe-compatible 6.0mm Digital Analog, REF...

PREAT Corporation / initiated 2024-06-25 / Open, Classified / root cause: manufacturing process

Preat Corp began a recall of Multiple Digital Analog Products labeled as: NobelBiocare Tri-Lobe-compatible 6.0mm Digital Analog, REF 9000703 Straumann Tissue-Level-compatible WN Digital... on . FDA classified it as Class II and gives the reason as "Due to a manufacturing issue, the product has a out of specification rotation of the digital analog used to visually inspect dental abutments and restorations". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2687-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
PREAT Corporation
Product
Multiple Digital Analog Products labeled as: NobelBiocare Tri-Lobe-compatible 6.0mm Digital Analog, REF 9000703 Straumann Tissue-Level-compatible WN Digital Analog, REF 9001303 3i Certain-compatible 5.0mm Digital Analog, REF 9001903 3i Certain-compatible 6.0mm Digital Analog, REF 9002003 Astra-compatible 3.0mm Digital Analog, REF 9002103 Astra-compatible Aqua 3.5/4.0mm Digital Analog, REF 9002203 Astra-compatible Lilac 4.5/5.0mm Digital Analog, REF 9002203 BioHorizons-compatible 3.0mm Digital Analog, REF 9005203 Legacy-compatible 3.0mm Digital Analog, REF 9005303 Digital Analog - BioHorizons - 3.5mm, Digital Analog - BioHorizons - 4.5mnm, REF 9005403 Legacy-compatible 3.5mm Digital Analog, REF 9005503 Legacy-compatible 5.7mm Digital Analog, REF 9008503 Astra EV 4.8mm Milled Titanium Abutment REF 9006767 Digital analogs are used for visual inspection of abutments and restoration in the dental laboratory.
Product code
NDP Accessories, Implant, Dental, Endosseous
Reason for recall
Due to a manufacturing issue, the product has a out of specification rotation of the digital analog used to visually inspect dental abutments and restorations.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On 07/22/2024, the firm sent an "URGENT: MEDICAL DEVICE RECALL UPDATE" Letter via certified mail informing customers that Digital Analogs were manufactured out of specification for the rotational feature which may result in the dental restoration/implant may be slightly different than the position of the actual implant in the patient's mouth. Customers are instructed to: Product in the scope of this recall should be discarded or returned to PREAT CORPORATION. For questions contact Quality Regulatory Affairs at 224-622-7191 or email [email protected]
Quantity in commerce
185 devices
Distribution
U.S. Nationwide distribution in the states of AK, AZ, FL, GA, IL, LA, MD, MI, MN, NY, OH, OK, OR, PA, RI, TX, UT, and WA
Product codes and lots
Model Numbers: 9000703, 9001303, 9001903, 9002003, 9002103, 9002203, 9002203, 9005203, 9005203, 9005303, 9005403, 9005503, 9008503, 9006767 UDI-DI codes: 00842092178690, 00842092181898, 00842092181959, 00842092178737, 00842092178775, 00842092178812, 00842092178812, 00842092178898, 00842092178898, 00842092178997, 00842092178928, 00842092178959, 00842092182130 Lot Numbers: 250219, 248890, 251432, 250602, 250418, 250213, 251459, 249071, 250335, 250223, 250336, 249510, 250215.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2687-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2687-2024Class IIOngoingVoluntary: Firm initiated