BD Kiestra Urine Culture Application Powered by BD Synapsys Informatics Solution; Catalog (Ref): 444900

Becton Dickinson & Co / initiated 2024-07-09 / Open, Classified

Becton Dickinson & Co. began a recall of BD Kiestra Urine Culture Application Powered by BD Synapsys Informatics Solution; Catalog (Ref): 444900 on . FDA classified it as Class II and gives the reason as "There is a potential risk that the image colony count does not adequately reflect the actual plate colony count and may lead to inaccurate counts that would impact patient diagnosis and subsequent care". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2585-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Becton Dickinson & Co
Product
BD Kiestra Urine Culture Application Powered by BD Synapsys Informatics Solution; Catalog (Ref): 444900;
Product code
KZB Automated Colony Counter
Reason for recall
There is a potential risk that the image colony count does not adequately reflect the actual plate colony count and may lead to inaccurate counts that would impact patient diagnosis and subsequent care.
FDA root cause
Under Investigation by firm
Action
On July 9, 2024, "URGENT: Medical Device Product Correction" letters were sent to customers. Actions for Clinical Users: Users must review the plate images against the actual agar plate to verify the automated colony count results. Actions to be taken by customer: 1. Ensure the contents of this notification are read and understood. 2. Share and post this customer letter with the users of this application at your facility and forward to any customers to whom you may have distributed the product to ensure awareness. 3. Complete the attached Customer Response Form and return to the BD contact noted on the form so that BD may acknowledge your receipt of this notification per FDA requirements. 4. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fisher s Lane, Rockville, MD 20852-9787 Contact Information: If you require further assistance, please contact: [email protected] or Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Complaints when prompted, Mon Fri 8:00am and 5:00pm CT or Email: [email protected].
Quantity in commerce
23 units
Distribution
US distribution to states of: CO, CT, IA, IL, KY, MD, MI, MN, NC, NJ, NY, OH, OK, SC, TN, TX and WA.
Product codes and lots
Catalog (Ref): 444900; UDI-DI: 00382904449001; Serial Numbers: IA00084, IA00054, IA00055, IA00088, IA00041, IA00071, IA00085, IA00007, IA00014, IA00005, IA00053, IA00073, IA00081, IA00011, IA00012, IA00027, SL01023, IA00026, IA00050, IA00034, IA00025, IA00086, IA00108.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2585-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2585-2024Class IIOngoingVoluntary: Firm initiated