Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2533-2024Class II

Verigene Gram Positive Blood Culture (BC-GP) Nucleic Acid Test

Luminex Corporation / initiated 2024-06-20 / Open, Classified

Luminex Corporation began a recall of Verigene Gram Positive Blood Culture (BC-GP) Nucleic Acid Test on . FDA classified it as Class II and gives the reason as "Potential for defective test cartridges which may result in false positive results for the Staphylococcus lugdunensis and Enterococcus faecium targets". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2533-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Luminex Corporation
Product
Verigene Gram Positive Blood Culture (BC-GP) Nucleic Acid Test
Product code
PAM Gram-Positive Bacteria And Their Resistance Markers
Reason for recall
Potential for defective test cartridges which may result in false positive results for the Staphylococcus lugdunensis and Enterococcus faecium targets.
FDA root cause
Under Investigation by firm
Action
Diasorin sent a recall notification on 06/20/2024, via email. Consignees were instructed to examine inventory and quarantine any affected units, arrange for replacement and confirm disposal of any of affected units on hand. Customers were asked to perform a retrospective review of results for VERIGENE Gram-Positive Blood Culture Nucleic Acid Test (BC-GP) to look for any false positives, distribute the notification to any relevant personnel, notify consignees if further distributed and complete and return the "Acknowledgement & Receipt Form."
Quantity in commerce
1,100 kits
Distribution
US Nationwide distribution including in the states of AZ, CA, CO, DE, FL, GA, IA, IL, KY, MD, MI, MN, MO, MS, NE, NV, OH, OK, SC, TN, TX, VA, VT, WA, WV.
Product codes and lots
GTIN Number: 00840487101575; Catalog Number: 20-006-018; Lot Number: 031824018A; Expiration Date: 09/16/2024

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2533-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2533-2024Class IIOngoingVoluntary: Firm initiated