Plum 360/A+ spare batteries. Item Number: SUB0000864
ICU Medical Inc began a recall of Plum 360/A+ spare batteries. Item Number: SUB0000864 on . FDA classified it as Class I and gives the reason as "Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2449-2024
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Abbott Medical
- Product
- Plum 360/A+ spare batteries. Item Number: SUB0000864.
- Product code
- FRN Pump, Infusion
- Reason for recall
- Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Consignees were notified on about 05/07/2024 via traceable mail with an "UPDATED URGENT MEDICAL DEVICE CORRECTION" letter to the original recall was expanding. Consignees were instructed to replace batteries the first time the pump issues a replace battery alarm. Continued use of pumps on battery after the first instance of the replace battery alarm will reduce the time from a low battery alarm to a depleted battery/infusion stopping. Additionally, pumps should be kept plugged into AC power, closely monitor if using battery power, have a backup pump available, and do not use a replacement battery with corroded battery terminals. Customers were also requested to ensure all users or potential users of these pumps are immediately made aware of this notification and proposed mitigations, and to complete and return the provided Response Form. For distributors, if the affected products have been potentially distributed, to immediately forward the notification to customers. For further inquiries, please contact ICU Medical using the following information: Global Complaint Management, 1-844-654-7780 (M-F, 8:00am 5:00pm CT) or [email protected] to report adverse events or product complaints; Technical Support 1-800-241-4002 (M-F, 8:00 am 6:00 pm CT) -additional information for technical assistance; or Field Corrections https://icumed.custhelp.com/app/market-action for Questions about this Field Correction Notice.
- Quantity in commerce
- 130,826 units
- Distribution
- Worldwide Distribution: US (nationwide) and OUS (foreign) countries of: Australia, Canada, France, Guadeloupe, Guatemala, Ireland, Italy, Oman, Paraguay, Qatar, Spain, Saudi Arabia, South Africa, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Uruguay.
- Product codes and lots
- Item Number: SUB0000864; Lot Numbers: 13620902, 13588166, 13651642, 13651688, 13651686, 13651685, 13492940, 13498252, 13498259, 13716418, 13716411, 13651680, 13620909, 13716514, 13588168, 13594285, 13620897, 13638726, 13620893, 13594289, 13638724, 13651671, 13594281, 13620898, 13659496, 13638723, 13594280, 13638728, 13588163, 13588169, 13588170, 13651640, 13638727, 13618384, 13769263, 13588165, 13588161, 13716427, 13570187, 13765276, 13651678, 13746712, 13749803, 13771023, 13771500, 13716336, 13588159, 13716329, 13594286, 13594287, 13620905, 13651677, 13594282.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2447-2024 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2448-2024 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2449-2024 | Class I | Ongoing | Voluntary: Firm initiated |