Disposable 5 Lead Cable and Lead Wire System, 50 inches, REF: LW-309DS50/5A
Stryker Sustainability Solutions began a recall of Disposable 5 Lead Cable and Lead Wire System, 50 inches, REF: LW-309DS50/5A on . FDA classified it as Class II and gives the reason as "Disposable 5 lead cable and lead wire system package that should contain lead systems, lead telemetry systems and a dual connect cable, may in fact contain, different ECG Leads, which are not approved for reprocessing...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2689-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Stryker Sustainability Solutions
- Product
- Disposable 5 Lead Cable and Lead Wire System, 50 inches, REF: LW-309DS50/5A
- Product code
- IKD Cable, Electrode
- Reason for recall
- Disposable 5 lead cable and lead wire system package that should contain lead systems, lead telemetry systems and a dual connect cable, may in fact contain, different ECG Leads, which are not approved for reprocessing, but that may have been reprocessed, which may lead to devices not performing as intended.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- On 7/2/24, recall notices were sent to customers informing them of the following: 1) Inspect affected devices for "TruLink" branding and if found, remove them from their point of use. 2) If any of the affected products have been forwarded to additional facilities (including but not limited to distribution warehouses), please contact these facilities and communicate this recall to them. Please instruct these facilities to complete the Business Reply Form. 3) Complete and return the business reply form via email to [email protected] Quality problems should be reported to the firm via phone +1-888-763-8803 or email http://www.stryker.com/productexperience
- Quantity in commerce
- 38,850
- Distribution
- US Nationwide distribution in the states of FL, WY, AZ, CA, LA, TX, MI, OR, CO, MS, TN, WI.
- Product codes and lots
- UDI-DI: 07613327483918, Lots: 0000111423, 0000115940, 0000117907, 0000124226, 0000129992, 0000169328, 0000176893, 0000178122, 0000181676, 0000206244, 0000210519, 0000211431, 0000213251, 0000217965, 0000219223, 0000221089, 0000228136, 0000231266, 0000236744, 0000238102, 0000240411, 0000240412, 0000250305, 0000291346, 0000356647, 0000372559, 0000445734, 0000474297, 0000541414, 0000549776, 11875844, 11893791, 11899068, 11902434, HY91020484, 11915419, 11925385, 11928534, 11931083, 11939652, 11947005, 11954851, 11963779, 11970640, 11977657, 11980331, 11984244, 11987875, 11991652, 11995462, 11995692, 12002389, 12010328, 12023953, 12026719, 12030554, 12039815, 12053079, 12056896, 12063693, 12080618, 12088836, 12093498, 12098299, 12102817, 12112224, 12148349, 12160590, 12170555, 12179425, 12185364, 12202282, 12204592, 12226747, 12238271, 12251638, 12253223, 12256926, 12260046, 12260098, 12271761, 12275519, 12277794, 12296434, 12327605, 12330999, 12351043, 12354510, 12355927, 12361109, 12363874, 12385821, 12386574, 12393911, 12401022, 12401392, 12409053, 12425416, 12431954, 12461855, 12469153, 12487497, 12490745, 12867127, 12879311, 12880822, 12892753, 12909629, 12939581, 12941498, 12959145, 12961324, 12963724, 12974181, 12998830, 13000131, 13006321, 13033322, 13051563, 13060599, 13067891, 13076246, 13092763, 13116447, 13121654, 13134031, 13146568, 13152147, 13158318, 13166734, 13203951, 13227500, 13242807, 13254161, 13280000, 13305113, 13325718, 12519983, 12532241, 12536829, 12541571, 12563008, 12564143, 12570286, 12581262, 12602792, 12607307, 12610241, 12616233, 12617893, 12647916, 12673690, 12678802, 12687516, 12693216, 12698775, 12701121, 12708984, 12735304, 12776269, 12777021, 12792610, 12795271, 12807688, 12809489, 12814655, 12824606, 12830261, 12851958, 12853941, 12867127, 12879311, 12880822, 12892753, 12909629, 12939581, 12941498, 12959145, 12961324, 12963724, 12974181, 12998830, 13000131, 13006321, 13033322, 13051563, 13060599, 13067891, 13076246, 13092763, 13116447, 13121654, 13134031, 13146568, 13152147, 13158318, 13166734, 13203951, 13227500, 13242807, 13254161, 13280000, 13305113, 13325718