Aisys CS2 with Et Control. This anesthesia gas machine is intended to provide general inhalation anesthesia...
Datex-Ohmeda, Inc. began a recall of Aisys CS2 with Et Control. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be... on . FDA classified it as Class I and gives the reason as "Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0083-2025
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- GE Healthcare
- Product
- Aisys CS2 with Et Control. This anesthesia gas machine is intended to provide general inhalation anesthesia and ventilatory support to patients and are to be used only by medical professionals trained and qualified in the administration of general anesthesia.
- Product code
- QSF Software Option For Anesthesia Gas Machine To Achieve And Maintain Targeted End Tidal Oxygen And Anesthetic Agents
- Reason for recall
- Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- GE Healthcare notified consignees on about 07/12/2024 via letter. Consignees were instructed to ensure all potential users complete the Preoperative Checkout procedure in its entirety, as detailed in the URM, make sure that the breathing circuit is correctly connected, only connect an auxiliary manual breathing circuit to the ACGO port, place the included addendum with the URM and inspect all potentially affected devices for the presence of covers applied to the ACGO Port and ACGO Switch. Additionally, customers were requested to complete and return the response form.
- Quantity in commerce
- 92 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Aland Islands, ALBANIA, Algeria, Argentina, Australia, Austria, Azerbaijan, BAHAMAS, Bahrain, Bangladesh, BARBADOS, BELARUS, Belgium, Benin, Bermuda, Bolivia, BONAIRE, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Cambodia, Cameroon, Canada, Chile, CHINA, Colombia, Costa Rica, Croatia, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, ETHIOPIA, Falkland Islands, Finland, France, French Polynesia, Gabon, Georgia, Germany, Ghana, Gilbrator, Greece, Guadeloupe, Guatemala, Guyana, Honduras, Hong Kong, Hungary, ICELAND, India, Indonesia, Iran, IRAQ, Ireland, Israel, Italy, Jamaica, JAPAN, Jordan, Kazakhstan, KENYA, Korea (Republic Of), Korea, Republic of, Kuwait, Laos, LATVIA, Lebanon, Libya, LITHUANIA, Macedonia, Malaysia, Maldives, Malta, MAURITIUS, Mexico, Moldova, Montenegro, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, North Macedonia, Norway, Oman, PAKISTAN, Palestine, State of, Panama, Paraguay, Peru, Philippines, Poland, Portugal, PUERTO RICO, Qatar, REUNION, ROMANIA, Russia, San Marino, Saudi Arabia, Senegal, Serbia, SINGAPORE, Slovakia, Slovenia, South Africa, Spain, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, THAILAND, Trinidad and Tobago, Tunisia, Turkey, T¿rkiye, Uganda, UKRAINE, United Arab Emirates, United Kingdom, Uruguay, UZBEKISTAN, Venezuela, Vietnam, Yemen.
- Product codes and lots
- GTIN: 00195278588128. Model Number: 1011-9055-000. All Serial Numbers. Not model or lot specific. All devices of the indicated products configured with the AGCO option are potentially affected.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0074-2025 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0075-2025 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0076-2025 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0077-2025 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0078-2025 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0079-2025 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0080-2025 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0081-2025 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0082-2025 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0083-2025 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0084-2025 | Class I | Ongoing | Voluntary: Firm initiated |