BD Pyxis QFill Replenishment Station, REF: 138904-01, medication cabinet system
CareFusion 303, Inc. began a recall of BD Pyxis QFill Replenishment Station, REF: 138904-01, medication cabinet system on . FDA classified it as Class II and gives the reason as "Due to software issue, there is a potential when a restock label is printed for a medication stored in a non-CUBIE location, that the incorrect bin for the medication is printed on the label". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2947-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- CareFusion 303, Inc
- Product
- BD Pyxis QFill Replenishment Station, REF: 138904-01, medication cabinet system
- Product code
- BRY Cabinet, Table And Tray, Anesthesia
- Reason for recall
- Due to software issue, there is a potential when a restock label is printed for a medication stored in a non-CUBIE location, that the incorrect bin for the medication is printed on the label.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On July 11, 2024, CareFusion a subsidiary of Becton Dickinson (BD) issued a "Urgent Medical Device Correction" Notification to affected customers via Email. BD also notified customers through customer software release notes beginning June 24, 2024. BD asked consignees to take the following actions: 1. When restocking the cabinet, refer to the MedBank restock workflow screen to verify the correct bin location rather than the printed label until the software update is provided. 2. Users should verify the contents of the bin location being restocked. Actions to be Taken by the Customer: 3.Please circulate this notice to all those who require awareness within your organization or to any organization whom you may have transferred BD Pyxis" MedBank" CUBIE" Replenishment Station devices to. 4.Please promptly complete and return the enclosed Customer Response Form within 15 days to acknowledge receipt of this notification and that you understand the information provided in this letter. 5. BD will perform remote upgrades to impacted customer devices with the updated CR/QFill station software. BD will contact customers by the end of August to initiate this scheduling.
- Quantity in commerce
- 3 units
- Distribution
- US Nationwide distribution in the states of CA, FL, KS, MN, NJ, NY, PA.
- Product codes and lots
- serial numbers: 44210411, 43880801, 44227426/ UDI: None
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2947-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2948-2024 | Class II | Ongoing | Voluntary: Firm initiated |