Z-3089-2024Class II

IBA Proton Therapy System - PROTEUS 235 ProteusPLUS and ProteusONE Software versions: PTS-8 versions, PTS-10...

Ion Beam Applications S.A. / initiated 2024-07-08 / Open, Classified

Ion Beam Applications S.A. began a recall of IBA Proton Therapy System - PROTEUS 235 ProteusPLUS and ProteusONE Software versions: PTS-8 versions, PTS-10 versions, PTS-11 versions, and PTS-12 versions on . FDA classified it as Class II and gives the reason as "Safety Parameters Verification Mechanisms can be deactivated in clinical runtime". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-3089-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Ion Beam Applications S.A.
Product
IBA Proton Therapy System - PROTEUS 235 ProteusPLUS and ProteusONE Software versions: PTS-8 versions, PTS-10 versions, PTS-11 versions, and PTS-12 versions
Product code
LHN System, Radiation Therapy, Charged-Particle, Medical
Reason for recall
Safety Parameters Verification Mechanisms can be deactivated in clinical runtime
FDA root cause
Under Investigation by firm
Action
IBA issued AMENDED Urgent Medical Device Correction letter (IBA REFERENCE NCIPT-27531) on 7/10/24 provided by IBA directly to the customer either by hand or by email. Letter states reason for recall, health risk and action to take: Regarding Proteus 235 with a PTS-8, PTS-10, PTS-11 or PTS-12 versions: actions: Regarding the malfunction of Safety Parameters Verification Mechanism within BAPP . IBA carried out an analysis to determine if the impacted sites have some safety parameters checked by the BAPP in their clinical site configuration with values leading to violations. The result of this analysis showed that there are no sites with safety parameters checked by the BAPP out of tolerance. Regarding the possibility to disable a Safety Parameters Verification Mechanism (concerns PTS-8, PTS-10, PTS-11 or PTS-12 versions): IBA will carry out an analysis to determine if the impacted sites have specific configuration parameters that allow to deactivate the Safety Parameters Verification Mechanisms in the clinical configuration and if the case, remove them from the clinical site configuration. The result of the analysis will be provided to you on July 31, 2024, at the latest. Final solutions: Regarding the malfunction of Safety Parameters Verification Mechanism within BAPP (only concerns PTS-10, PTS-11 or PTS-12 versions): IBA will ensure that the system cannot be used in clinical with safety parameters checked by the BAPP out of tolerances. Regarding the possibility to disable a Safety Parameters Verification Mechanism (concerns PTS-8, PTS-10, PTS-11 or PTS-12 versions): IBA will ensure that the system displays a Red Banner informing the customer that the system must not be used for clinical activities when safety parameter Verification Mechanisms are deactivated to prevent the users from treating patients. The intermediate action will be implemented for your site by August 31, 2024, at the latest. The final solution will be implemented for your site by June 2026. Fi
Quantity in commerce
35 units
Distribution
Worldwide - US Nationwide distribution in the states of AK, NJ, PA, VA,OK,FL, KA, IL, LA, MI, TN , TX, VA and the countries of BE, CZ, DE, ES, FR, IN, IT, JP, NL, PL, RU, SE, SG, TW.
Product codes and lots
UDI-DI: 05404013801138 Serial Numbers: PAT.107 (DE), PAT.108 (US), PAT.109 (US), PAT.110 (US), PAT.111 (FR), PAT.112 (US), PAT.113 (US), PAT.114 (CZ), PAT.115 (IT), SAT.116 (US), SAT.117 (PL), SAT.118 (RU), SAT.119 (US), SAT.120 (DE), SAT.122 (SE), SAT.123 (US), SAT.125 (IN), SAT.126 (US), SAT.127 (TW), SAT.132 (NL), SAT.133 (US), SAT.136 (IN), SAT.140 (US), SBF.101 (FR), SBF.103 (JP), SBF.104 (JP), SBF.105 (US), SBF.107 (FR), SBF.112 (BE), SBF.113 (US), SBF.117 (ES), SBF.124 (IT), SBF.125 (SG), SBF.128 (US), SBF.135 (US).

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3089-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3089-2024Class IIOngoingVoluntary: Firm initiated