Z-2715-2024Class II

GM60A. Digital Diagnostic Mobile X-Ray System

Samsung HME America, Inc. / initiated 2024-07-08 / Open, Classified / root cause: design

NeuroLogica Corporation began a recall of GM60A. Digital Diagnostic Mobile X-Ray System on . FDA classified it as Class II and gives the reason as "Potential for the anti-fall system of the device arm to fail". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-2715-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Samsung HME America, Inc.
Product
GM60A. Digital Diagnostic Mobile X-Ray System.
Product code
IZL System, X-Ray, Mobile
Reason for recall
Potential for the anti-fall system of the device arm to fail.
Root cause tag
design
FDA root cause
Device Design
Action
Consignees were emailed an "Urgent -- Medical Device Correction; GM60A Digital X-ray Imaging System" notification the week of 7/8/2024. The notification instructs consignees to share the recall notification with those in their organization that utilize the affected devices and forward the notification if devices have been transferred. Affected systems may continue to be used. Consignees are to check on-hand devices to confirm if they are affected by the field action and return the response form back indicating any devices that require correction by email to [email protected] within 10 business days. Samsung Electronics will send service engineers to affected consignees to correct devices free of charge. Service will be coordinated by Samsung Electronics and should be completed by October 2024. Consignees with any questions are to contact their local Samsung Electronics representative.
Quantity in commerce
31 units
Distribution
US Nationwide distribution in the states of AR, CT, GA, IN, KS, MA, MD, MO, MS, NC, ND, NJ, NV, OH, PA, VA, WV & DC.
Product codes and lots
UDI: 08806088511153/ 08806088088266; Serial No. G69812, G69821, G69824, G69986, G69987, G71422, G71425, G71431, G71434, G71795, G71798, G71802, G72052, G72053, G72264, G72279, G72296, G72301, G72310, G72313, G72319, G72320, G72481, G72484, G72595, G72601, G72853, G72875, G73179, G73184, G73542; Samsung Serial No. 50WFM9BG900003M, 50WFM9BG900004J, 50WFM9BG900005H, 50WFM9BGA00001T, 50WFM9BGA00002B, 50WFM9BH300001E, 50WFM9BH300002W, 50WFM9BH300003Z, 50WFM9BH300004N, 50WFM9BH400001X, 50WFM9BH400002W, 50WFM9BH400003A, 50WFM9BH500001H, 50WFM9BH500002Y, 50WFM9BH600001P, 50WFM9BH600002L, 50WFM9BH600003F, 50WFM9BH600004Y, 50WFM9BH600005W, 50WFM9BH600006N, 50WFM9BH600007K, 50WFM9BH600008J, 50WFM9BH700002B, 50WFM9BH700001P, 50WFM9BH700003M, 50WFM9BH700004E, 50WFM9BH800002E, 50WFM9BH800003D, 50WFM9BH900001X, 50WFM9BH900002D, 50WFM9BHB00001T.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2715-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2715-2024Class IIOngoingVoluntary: Firm initiated