Z-2738-2024Class II
DR 800. Digital Radiography X-ray System
Agfa N.V. / initiated 2024-07-29 / Open, Classified / root cause: design
Agfa N.V. began a recall of DR 800. Digital Radiography X-ray System on . FDA classified it as Class II and gives the reason as "Potential for the front lever chain of the DR 800 table to fail". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2738-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Agfa N.V.
- Product
- DR 800. Digital Radiography X-ray System.
- Product code
- JAA System, X-Ray, Fluoroscopic, Image-Intensified
- Reason for recall
- Potential for the front lever chain of the DR 800 table to fail.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Consignees were sent an URGENT FIELD SAFETY NOTICE by email, dated 7/29/24. In the notice, consignees are asked to check the correct tightening of the floor plate fixing bolts if a high and persistent noise is heard from the device during the tilting or elevation movement of the tabletop. If the noise persists, the device must be stopped immediately in a horizontal position at the maximum height. Agfa will coordinate with consignees to schedule an inspection of affected devices and to install hardware intended to reinforce equipment before 10/15/24. The recall notification is to be shared within consignee organizations and forwarded to those devices were transferred to. Consignees are asked to maintain awareness of this notice until devices are corrected. Completed response forms are to be provided via email to [email protected]. Consignees with any questions are to call 1-877-777-2432 and reference PRB2001106.
- Quantity in commerce
- 141 units
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, FL, IL, IN, KS, LA, MA, MO, NJ, NC, OH, OK, SC, VA, WV, WI; and the countries of AU, BE, BW, CA, CO, CZ, DE, ES, FR, GB, GR, IL, IN, IT, KZ, MX, PL, RS, UZ, ZA.
- Product codes and lots
- Type No. 6010/200; UDI-DI: 05414904251805; Device Serial No. 925, 945, 948, 949, 990, 1003, 1022, 1116, 1136, 1161, 1229, 1239, 1401, 1404, 1453, 1460, 1489, 1506, 1532, 1568, 1704, 1733, 1757, 1888, 1893, 1950, 1954, 1957, 2010, 2070, 2130, 2244, 2297, 2371, 2378, 2901, 2931, 3068, 3217, 1507, 643, 763, 903, 915, 930, 969, 984, 992, 1032, 1042, 1049, 1052, 1123, 1148, 1197, 1217, 1323, 1341, 1203, 1279, 1458, 1509, 1521, 1627, 1676, 1691, 1756, 1828, 1850, 1878, 1921, 1924, 1926, 1932, 1938, 1941, 1943, 1956, 2079, 2088, 2094, 2101, 2115, 2129, 2131, 2157, 2169, 2192, 2193, 2225, 2233, 2243, 2251, 2253, 2254, 2277, 2289, 2310, 2314, 2318, 2333, 2340, 2359, 2377, 2409, 2415, 2416, 2428, 2436, 2437, 2449, 2450, 2475, 2483, 2513, 2538, 2580, 2642, 2678, 2687, 2756, 2781, 2833, 2842, 2873, 2886, 2892, 2941, 2955, 2992, 3101, 3134, 3143, 3151, 3161, 1920, 3006, 3040; Table Serial No. 32497, 34673, 34695, 34713, 34717, 34976, 34990, 35350, 36241, 36267, 36462, 37150, 37177, 38592, 38595, 38624, 38648, 39729, 39763, 39773, 39796, 41160, 41179, 41192, 42084, 42088, 42139, 42138, 43208, 43239, 43250, 44203, 45270, 45274, 46108, 46130, 49894, 51360, 52275, 53884, 39755, 31908, 32486, 32977, 34257, 34353, 34355, 34678, 34957, 34970, 34979, 35365, 35371, 38591, 35756, 36246, 36442, 36883, 37137, 37869, 37909, 36889, 37221, 38630, 39739, 39757, 40634, 41132, 41147, 41191, 42066, 42071, 42076, 42098, 42101, 42102, 42113, 42115, 42117, 42118, 42140, 43259, 43265, 43269, 43288, 44180, 44188, 44192, 44212, 44233, 44253, 44254, 45219, 45227, 45235, 45259, 45249, 45264, 45260, 45269, 45282, 45283, 45285, 45299, 45304, 46076, 46113, 46145, 46148, 46156, 46166, 46167, 47188, 46175, 47186, 47204, 47207, 47227, 47241, 47273, 48349, 48361, 48376, 48413, 49808, 49830, 49834, 49869, 49882, 49887, 51364, 51381, 51422, 52303, 52322, 52327, 52333, 52337, 42097, 51428, 52255.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2738-2024 | Class II | Ongoing | Voluntary: Firm initiated |