Z-3065-2024Class II

Dimension Vista Total Bilirubin (TBIL) Flex Reagent Cartridge; Siemens Material Number (SMN): 10445146

Siemens Healthcare Diagnostics Inc. / initiated 2024-07-30 / Open, Classified / root cause: software

Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Vista Total Bilirubin (TBIL) Flex Reagent Cartridge; Siemens Material Number (SMN): 10445146 on . FDA classified it as Class II and gives the reason as "During manufacturing, the last three tests from the 160 test flex of lot 23206BA (wells 8 or 10) were processed on the Dimension Vista platform. There was an issue with the filling line which led to some wells to...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-3065-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Dimension Vista Total Bilirubin (TBIL) Flex Reagent Cartridge; Siemens Material Number (SMN): 10445146;
Product code
CIG Diazo Colorimetry, Bilirubin
Reason for recall
During manufacturing, the last three tests from the 160 test flex of lot 23206BA (wells 8 or 10) were processed on the Dimension Vista platform. There was an issue with the filling line which led to some wells to underfill in lot 23206BA. Calibrator, quality control (QC) and patient results could have been impacted.
Root cause tag
software
FDA root cause
Software Manufacturing/Software Deployment
Action
On July 30, 2024 URGENT MEDICAL DEVICE CORRECTION letters were sent to consignees. Impact to Results: 1. Erroneously depressed or elevated total bilirubin patient results may have occurred, even if QC was within range. 2. A negative bias of >-98% (erroneous result <0.1 mg/dL [2 mol/L] at a true concentration of 4.0 mg/dL [69 mol/L]) and a positive bias of up to 3,950% (erroneous result 16.2 mg/dL [277 mol/L] at a true concentration of 0.4 mg/dL [7 mol/L]) was observed. Results of this test should always be interpreted in conjunction with the patient s medical history, clinical presentation and other findings. Customer Actions: 1. Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. 2. Complete and return the Field Correction Effectiveness Check form attached within 30 days and indicate product replacement needs if retesting of patients is performed. 3. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Resolution: The cause of the filling issue has been determined and resolved by Siemens Healthineers. Mitigation that will prevent this issue from occurring in the future has been put in place. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
Quantity in commerce
5,371 units (4,394 US, 977 OUS)
Distribution
Domestic: Nationwide Distribution. International: Austria, Belgium, Canada, Czech Republic, Denmark, France, Germany, Italy, Qatar, Slovenia, Spain, Switzerland.
Product codes and lots
Siemens Material Number (SMN): 10445146; UDI-DI: 00842768015717; Lot Number: 23206BA;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3065-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3065-2024Class IIOngoingVoluntary: Firm initiated