BioPince Ultra Full Core Biopsy Instrument: REF: REF: 360-1080-01, 360-1080-02, 360-1080-03, 360-1580-01...
Argon Medical Devices, Inc began a recall of BioPince Ultra Full Core Biopsy Instrument: REF: REF: 360-1080-01, 360-1080-02, 360-1080-03, 360-1580-01, 360-1580-02, 360-1580-03, 360-2080-01, 360-2080-02... on . FDA classified it as Class II and gives the reason as "Holes have been identified in biopsy instrument packaging trays compromising the sterile barrier, which may lead to cross contamination, or the introduction of micro-organisms into the access site, which could lead to...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-3192-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Argon Medical Devices, Inc
- Product
- BioPince Ultra Full Core Biopsy Instrument: REF: REF: 360-1080-01, 360-1080-02, 360-1080-03, 360-1580-01, 360-1580-02, 360-1580-03, 360-2080-01, 360-2080-02, 360-2080-03, 370-1080-01, 370-1080-02, 370-1080-03, 370-1580-01, 370-1580-02, 370-1580-03; Tru-Core II Biopsy Instrument, REF: 763114100x, 763116100X, 763116160X, 763118100X, 763118200X, 763120100X, 763120160X, 763418200X, 763418250X, 763114200X
- Product code
- KNW Instrument, Biopsy
- Reason for recall
- Holes have been identified in biopsy instrument packaging trays compromising the sterile barrier, which may lead to cross contamination, or the introduction of micro-organisms into the access site, which could lead to an infectious process, bacteremia, or sepsis.
- Root cause tag
- packaging
- FDA root cause
- Packaging change control
- Action
- On 7/2/24, recall notices were mailed and emailed to Physician, Clinician, or Hospital Administrators who were asked to do the following: 1) Discontinue use of affected devices. 2) Segregate, quarantine, and return affected devices to the recalling firm. 3) Share this recall notice with the following parties to ensure your consignees receive a copy of this notice and the associated response form: a. Anyone who needs to be aware within your organization b. Any organization where the potentially affected products have been transferred or further distributed c. Anyone within your organization who may use this device 4) Complete and return the response form via email to [email protected]
- Quantity in commerce
- 95,430
- Distribution
- Worldwide - US Nationwide distribution including in the states of AZ, FL, OH, NY, MO, CO, IL, NC, CA, AK, PA, MN, ND, TX, SC, AR, WI, OK, TN, KS, UT, NJ, VA, MA, KY, IN, AL, DE, MT, NE, WV, NV, HI, OR, WA, NH, LA, CT, MI, GA, MD, ID, MS, IA, RI, NM, VT and the countries of ITA, TUN, ROU, LTU, BEL, GBR, CAN, TUR, POL, PAN, THA, PRT, TWN, AUT, NZL, AUS, BGR, MEX, CHE, NLD, PHL, PER, FRA, SGP, ZAF, CZE, EST, ESP, OMN, NPL, COL, PRI, ISR, JPN, KEN, BHR, ARE, CHN.
- Product codes and lots
- REF/Box UDI-DI/Unit UDI-DI/Lot(Expiration Date): 360-1080-01/20886333004373/00886333004379/11563436(22-Mar-2027), 11568010(16-Apr-2027), 11568790(1-May-2027), 11569552(7-May-2027), 11572560(20-May-2027) 360-1080-02/20886333224009/00886333224005/11563678(22-Mar-27), 11562475(05-Mar-27), 11562476(14-Mar-27), 11566932(08-Apr-27), 11567139(11-Apr-27), 11568361(16-Apr-27), 11569013(23-Apr-27), 11569844(29-Apr-27), 11571094(02-May-27), 11572044(08-May-27), 11572886(26-May-27); 360-1080-03/20886333224016/00886333224012/11569845(22-Apr-27), 11562477(04-Mar-27), 11564861(26-Mar-27), 11567888(12-Apr-27), 11568554(18-Apr-27), 11570075(04-May-27); 360-1580-01/20886333004359/00886333004355/11563438(21-Mar-2027), 11563439(19-Mar-2027); 360-1580-02/20886333224023/00886333224029/11562478(14-Mar-27), 11563440(18-Mar-27), 11563682(20-Mar-27), 11565238(03-Apr-27), 11566882(09-Apr-27), 11570132(01-May-27), 11570814(09-May-27), 11573339(20-May-27); 360-1580-03/20886333224030/00886333224036/11570288(02-May-27), 11563442(20-Mar-27), 11566173(03-Apr-27), 11566883(10-Apr-27), 11568153(16-Apr-27), 11570813(08-May-27); 360-2080-01/20886333004342/00886333004348/11563564(26-Mar-2027), 11566036(8-Apr-2027), 11567992(16-Apr-2027), 11570142(1-May-27), 11570863(8-May-27); 360-2080-02/20886333224047/00886333224043/11562964(07-Mar-27), 11563922(15-Mar-27), 11564189(22-Mar-27); 360-2080-03/20886333224054/00886333224050/11563443(15-Mar-27), 11566936(11-Apr-27), 11567889(22-Apr-2027); 370-1080-01/20886333004380/00886333004386/11564795(26-Mar-27); 370-1080-02/20886333224061/00886333224067/11562608(05-Mar-27), 11562748(04-Mar-27), 11564843(26-Mar-27), 11566037(08-Apr-27), 11571101(30-Apr-27), 11571360(06-May-27); 370-1080-03/20886333224078/00886333224074/11564844(22-Mar-27), 11566690(16-Apr-27), 11571100(01-May-27), 11573030(13-May-27); 370-1580-01/20886333004366/00886333004382/11564366(26-Mar-27); 370-1580-02/20886333224085/00886333224081/11571824(06-May-27), 11562114(07-Mar-27), 11566038(08-Apr-27), 11571824(06-May-27); 370-1580-03/20886333224092/00886333224098/11569414(22-Apr-27), 11571104(01-May-27); 763114100x/20886333006643/00886333006649/11563673(11-May-2029), 11570356(23-Apr-2029), 11571058(4-Apr-2029), 11572456(29-May-2029), 11572868(13-May-2029), 11573300(22-May-2029); 763116100X/20886333006612/00886333006618/11562815(7-Mar-2029), 11571061(29-Apr-2029); 763116160X/20886333006605/00886333006601/11571063(1-May-2029); 763118100X/20886333006582/00886333006588/11564330(26-Mar-2029); 763118200X/20886333006650/00886333006656/11562085(5-Mar-2029), 11571065(4-APR-2029); 763120100X/20886333006568/00886333006564/11564331(26-Mar-2029); 763120160X/20886333006575/00886333006571/11565618(10-Apr-2029); 763418200X/20886333006803/00886333006809/11566213(10-Apr-2029); 763418250X/20886333006810/00886333006816/11564594(22-Mar-2029), 11565619(4-Apr-2029), 11572175(22-May-2029); 763114200X/20886333006629/00886333006625/11566106(4-Apr-2029)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3192-2024 | Class II | Ongoing | Voluntary: Firm initiated |