Z-3192-2024Class II

BioPince Ultra Full Core Biopsy Instrument: REF: REF: 360-1080-01, 360-1080-02, 360-1080-03, 360-1580-01...

Argon Medical Devices, Inc / initiated 2024-07-02 / Open, Classified / root cause: packaging

Argon Medical Devices, Inc began a recall of BioPince Ultra Full Core Biopsy Instrument: REF: REF: 360-1080-01, 360-1080-02, 360-1080-03, 360-1580-01, 360-1580-02, 360-1580-03, 360-2080-01, 360-2080-02... on . FDA classified it as Class II and gives the reason as "Holes have been identified in biopsy instrument packaging trays compromising the sterile barrier, which may lead to cross contamination, or the introduction of micro-organisms into the access site, which could lead to...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-3192-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Argon Medical Devices, Inc
Product
BioPince Ultra Full Core Biopsy Instrument: REF: REF: 360-1080-01, 360-1080-02, 360-1080-03, 360-1580-01, 360-1580-02, 360-1580-03, 360-2080-01, 360-2080-02, 360-2080-03, 370-1080-01, 370-1080-02, 370-1080-03, 370-1580-01, 370-1580-02, 370-1580-03; Tru-Core II Biopsy Instrument, REF: 763114100x, 763116100X, 763116160X, 763118100X, 763118200X, 763120100X, 763120160X, 763418200X, 763418250X, 763114200X
Product code
KNW Instrument, Biopsy
Reason for recall
Holes have been identified in biopsy instrument packaging trays compromising the sterile barrier, which may lead to cross contamination, or the introduction of micro-organisms into the access site, which could lead to an infectious process, bacteremia, or sepsis.
Root cause tag
packaging
FDA root cause
Packaging change control
Action
On 7/2/24, recall notices were mailed and emailed to Physician, Clinician, or Hospital Administrators who were asked to do the following: 1) Discontinue use of affected devices. 2) Segregate, quarantine, and return affected devices to the recalling firm. 3) Share this recall notice with the following parties to ensure your consignees receive a copy of this notice and the associated response form: a. Anyone who needs to be aware within your organization b. Any organization where the potentially affected products have been transferred or further distributed c. Anyone within your organization who may use this device 4) Complete and return the response form via email to [email protected]
Quantity in commerce
95,430
Distribution
Worldwide - US Nationwide distribution including in the states of AZ, FL, OH, NY, MO, CO, IL, NC, CA, AK, PA, MN, ND, TX, SC, AR, WI, OK, TN, KS, UT, NJ, VA, MA, KY, IN, AL, DE, MT, NE, WV, NV, HI, OR, WA, NH, LA, CT, MI, GA, MD, ID, MS, IA, RI, NM, VT and the countries of ITA, TUN, ROU, LTU, BEL, GBR, CAN, TUR, POL, PAN, THA, PRT, TWN, AUT, NZL, AUS, BGR, MEX, CHE, NLD, PHL, PER, FRA, SGP, ZAF, CZE, EST, ESP, OMN, NPL, COL, PRI, ISR, JPN, KEN, BHR, ARE, CHN.
Product codes and lots
REF/Box UDI-DI/Unit UDI-DI/Lot(Expiration Date): 360-1080-01/20886333004373/00886333004379/11563436(22-Mar-2027), 11568010(16-Apr-2027), 11568790(1-May-2027), 11569552(7-May-2027), 11572560(20-May-2027) 360-1080-02/20886333224009/00886333224005/11563678(22-Mar-27), 11562475(05-Mar-27), 11562476(14-Mar-27), 11566932(08-Apr-27), 11567139(11-Apr-27), 11568361(16-Apr-27), 11569013(23-Apr-27), 11569844(29-Apr-27), 11571094(02-May-27), 11572044(08-May-27), 11572886(26-May-27); 360-1080-03/20886333224016/00886333224012/11569845(22-Apr-27), 11562477(04-Mar-27), 11564861(26-Mar-27), 11567888(12-Apr-27), 11568554(18-Apr-27), 11570075(04-May-27); 360-1580-01/20886333004359/00886333004355/11563438(21-Mar-2027), 11563439(19-Mar-2027); 360-1580-02/20886333224023/00886333224029/11562478(14-Mar-27), 11563440(18-Mar-27), 11563682(20-Mar-27), 11565238(03-Apr-27), 11566882(09-Apr-27), 11570132(01-May-27), 11570814(09-May-27), 11573339(20-May-27); 360-1580-03/20886333224030/00886333224036/11570288(02-May-27), 11563442(20-Mar-27), 11566173(03-Apr-27), 11566883(10-Apr-27), 11568153(16-Apr-27), 11570813(08-May-27); 360-2080-01/20886333004342/00886333004348/11563564(26-Mar-2027), 11566036(8-Apr-2027), 11567992(16-Apr-2027), 11570142(1-May-27), 11570863(8-May-27); 360-2080-02/20886333224047/00886333224043/11562964(07-Mar-27), 11563922(15-Mar-27), 11564189(22-Mar-27); 360-2080-03/20886333224054/00886333224050/11563443(15-Mar-27), 11566936(11-Apr-27), 11567889(22-Apr-2027); 370-1080-01/20886333004380/00886333004386/11564795(26-Mar-27); 370-1080-02/20886333224061/00886333224067/11562608(05-Mar-27), 11562748(04-Mar-27), 11564843(26-Mar-27), 11566037(08-Apr-27), 11571101(30-Apr-27), 11571360(06-May-27); 370-1080-03/20886333224078/00886333224074/11564844(22-Mar-27), 11566690(16-Apr-27), 11571100(01-May-27), 11573030(13-May-27); 370-1580-01/20886333004366/00886333004382/11564366(26-Mar-27); 370-1580-02/20886333224085/00886333224081/11571824(06-May-27), 11562114(07-Mar-27), 11566038(08-Apr-27), 11571824(06-May-27); 370-1580-03/20886333224092/00886333224098/11569414(22-Apr-27), 11571104(01-May-27); 763114100x/20886333006643/00886333006649/11563673(11-May-2029), 11570356(23-Apr-2029), 11571058(4-Apr-2029), 11572456(29-May-2029), 11572868(13-May-2029), 11573300(22-May-2029); 763116100X/20886333006612/00886333006618/11562815(7-Mar-2029), 11571061(29-Apr-2029); 763116160X/20886333006605/00886333006601/11571063(1-May-2029); 763118100X/20886333006582/00886333006588/11564330(26-Mar-2029); 763118200X/20886333006650/00886333006656/11562085(5-Mar-2029), 11571065(4-APR-2029); 763120100X/20886333006568/00886333006564/11564331(26-Mar-2029); 763120160X/20886333006575/00886333006571/11565618(10-Apr-2029); 763418200X/20886333006803/00886333006809/11566213(10-Apr-2029); 763418250X/20886333006810/00886333006816/11564594(22-Mar-2029), 11565619(4-Apr-2029), 11572175(22-May-2029); 763114200X/20886333006629/00886333006625/11566106(4-Apr-2029)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3192-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3192-2024Class IIOngoingVoluntary: Firm initiated