Medline convenience kits containing semi-rigid suction liners labeled as follows: a) OPEN HEART KIT,ALTA...
MEDLINE INDUSTRIES, LP - Northfield began a recall of Medline convenience kits containing semi-rigid suction liners labeled as follows: a) OPEN HEART KIT,ALTA BATES HOSP, Pack Number DYK1002174I; b) ENDO KIT, Pack... on . FDA classified it as Class II and gives the reason as "Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-3107-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Medline convenience kits containing semi-rigid suction liners labeled as follows: a) OPEN HEART KIT,ALTA BATES HOSP, Pack Number DYK1002174I; b) ENDO KIT, Pack Number DYKE1558B; c) ENDO KIT, Pack Number DYKE1804A; d) VASCULAR-LF, Pack Number DYNJ900714C; e) MAJOR CARDIOVASC ADULT, Pack Number DYNJ905472B; f) AV FISTULA, Pack Number DYNJ908096A
- Product code
- OEZ Cardiovascular Procedure Kit
- Reason for recall
- Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Medline Industries issued a MEDICAL DEVICE RECALL notice to its consignees on 06/27/2024 via US mail and email. The notice explained the problem with the device, potential risk to the patient, and requested the consignee immediately check their stock for the affected item number and the affected lot numbers and destroy any affected inventory. Distributors were directed to notify their customers of the recall and request the affected products be destroyed. For questions: contact the Recall Department at 866-359-1704 or [email protected].
- Quantity in commerce
- 1,646 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Panama.
- Product codes and lots
- a) DYK1002174I: UDI/DI case 40195327332403, UDI/DI each 10195327332402; Lot Numbers: 24CLA959; b) DYKE1558B: UDI/DI case 40195327258512, UDI/DI each 10195327258511; Lot Numbers: 24DMA783, 24BMG878, 23LMG790; c) DYKE1804A: UDI/DI case 40195327023035, UDI/DI each 10195327023034; Lot Numbers: 24BMI437, 24AME311, 23LMC958; d) DYNJ900714C: UDI/DI case 40193489951364, UDI/DI each 10193489951363; Lot Numbers: 24FBC609; e) DYNJ905472B: UDI/DI case 40195327200573, UDI/DI each 10195327200572; Lot Numbers: 24BBS291; f) DYNJ908096A: UDI/DI case 40195327507061, UDI/DI each 10195327507060; Lot Numbers: 24ABX948, 24ABB080
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3105-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3106-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3107-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3108-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3109-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3110-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3111-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3112-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3113-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3114-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3115-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3116-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3117-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3118-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3119-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3120-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3121-2024 | Class II | Ongoing | Voluntary: Firm initiated |