Z-3073-2024Class II

Medtronic Handset with Communicator, Model TH90Q01, InterStim Smart Programmer and Communicator Kits...

Medtronic Neuromodulation / initiated 2024-06-17 / Open, Classified / root cause: software

Medtronic Neuromodulation began a recall of Medtronic Handset with Communicator, Model TH90Q01, InterStim Smart Programmer and Communicator Kits, containing the Model HH90 Handset and the TM90... on . FDA classified it as Class II and gives the reason as "A subset of the handsets within the kit may not be able to complete the pairing process with the communicator upon initial setup". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-3073-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic Neuromodulation
Product
Medtronic Handset with Communicator, Model TH90Q01, InterStim Smart Programmer and Communicator Kits, containing the Model HH90 Handset and the TM90 Communicator.
Product code
EZW Stimulator, Electrical, Implantable, For Incontinence
Reason for recall
A subset of the handsets within the kit may not be able to complete the pairing process with the communicator upon initial setup.
Root cause tag
software
FDA root cause
Software change control
Action
The recalling firm issued letters dated June 2024 beginning 6/17/2024 to the Medtronic Field Service Representatives to notify consignees and retrieve a subset of the InterStim Smart Programmer and Communicator Kits (TH90Q01). The notification informed the Field Service Representatives of the Field Corrective Action and provided instruction on notifying the consignees and retrieving the device(s). Beginning on June 17, 2024, US initiated communications and notifications dated June 2024 were physically hand-delivered to each consignee with the impacted device(s). The letter described the issue and provided Required Customer Actions, which included completing and signing the Customer Confirmation form acknowledging receipt of the notification and sharing the notice with all that needed to be aware. An appendix was enclosed which listed the serial numbers of the affected products.
Quantity in commerce
211 devices
Distribution
US Nationwide distribution.
Product codes and lots
GTIN 0076300092754, serial numbers: NPV0082351, NPV0085575, NPV0082308, NPV0082309, NPV0082314, NPV0082328, NPV0082329, NPV0082349, NPV0082412, NPV0082483, NPV0082494, NPV0082608, NPV0082631, NPV0082646, NPV0082684, NPV0082689, NPV0082690, NPV0082691, NPV0082756, NPV0082866, NPV0082869, NPV0082899, NPV0082921, NPV0083001, NPV0083071, NPV0083084, NPV0083116, NPV0083125, NPV0083171, NPV0083507, NPV0083577, NPV0083629, NPV0083632, NPV0083853, NPV0083883, NPV0083917, NPV0083944, NPV0083948, NPV0083954, NPV0083972, NPV0084077, NPV0084084, NPV0084104, NPV0084113, NPV0084119, NPV0084125, NPV0084177, NPV0084211, NPV0084253, NPV0084283, NPV0084293, NPV0084295, NPV0084304, NPV0084305, NPV0084311, NPV0084312, NPV0084313, NPV0084317, NPV0084330, NPV0084340, NPV0084353, NPV0084355, NPV0084428, NPV0084473, NPV0084523, NPV0084610, NPV0084634, NPV0084679, NPV0084681, NPV0084695, NPV0084697, NPV0084751, NPV0084753, NPV0084754, NPV0084767, NPV0085004, NPV0085034, NPV0085036, NPV0085057, NPV0085059, NPV0085077, NPV0085081, NPV0085473, NPV0085510, NPV0085529, NPV0085580, NPV0085601, NPV0085616, NPV0085626, NPV0085631, NPV0085647, NPV0085653, NPV0082276, NPV0082278, NPV0082281, NPV0082304, NPV0082313, NPV0082316, NPV0082320, NPV0082340, NPV0082352, NPV0082358, NPV0082376, NPV0082377, NPV0082418, NPV0082421, NPV0082431, NPV0082450, NPV0082451, NPV0082476, NPV0082480, NPV0082508, NPV0082513, NPV0082514, NPV0082520, NPV0082524, NPV0082537, NPV0082539, NPV0082543, NPV0082559, NPV0082562, NPV0082569, NPV0082571, NPV0082572, NPV0082591, NPV0082635, NPV0082666, NPV0082674, NPV0082695, NPV0082701, NPV0082722, NPV0082723, NPV0082740, NPV0082769, NPV0082779, NPV0082782, NPV0082789, NPV0082806, NPV0082859, NPV0082871, NPV0082879, NPV0082882, NPV0082897, NPV0082913, NPV0082915, NPV0082916, NPV0082923, NPV0082928, NPV0082931, NPV0082944, NPV0082945, NPV0082949, NPV0082952, NPV0082960, NPV0082963, NPV0083005, NPV0083027, NPV0083036, NPV0083047, NPV0083052, NPV0083053, NPV0083104, NPV0083129, NPV0083134, NPV0083175, NPV0083178, NPV0083189, NPV0083192, NPV0083200, NPV0083225, NPV0083228, NPV0083229, NPV0083230, NPV0083237, NPV0083277, NPV0083289, NPV0083302, NPV0083312, NPV0083329, NPV0083332, NPV0083344, NPV0083345, NPV0083349, NPV0083371, NPV0083396, NPV0083484, NPV0083506, NPV0083522, NPV0083526, NPV0083555, NPV0083575, NPV0083596, NPV0083604, NPV0083613, NPV0083618, NPV0083646, NPV0083649, NPV0083688, NPV0083708, NPV0083712, NPV0083716, NPV0083729, NPV0083738, NPV0083827, NPV0083842, NPV0083845, NPV0083846, NPV0083847, NPV0083859, NPV0083937, NPV0084126.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3073-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3073-2024Class IIOngoingVoluntary: Firm initiated