CARDIOSAVE RESCUE INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate...

Datascope Corp. / initiated 2024-08-08 / Open, Classified / root cause: software

Datascope Corp. began a recall of CARDIOSAVE RESCUE INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons on . FDA classified it as Class I and gives the reason as "Firm has developed a software correction (Revision D.00) outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-3135-2024
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Datascope Corp.
Product
CARDIOSAVE RESCUE INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.
Product code
DSP System, Balloon, Intra-Aortic And Control
Reason for recall
Firm has developed a software correction (Revision D.00) outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and Cardiosave Rescue Intra-Aortic Balloon Pump devices.
Root cause tag
software
FDA root cause
Software design
Action
Consignees were mailed an URGENT FIELD SAFETY NOTICE dated 8/5/2024. This notification informs consignees of the deployment of previously installed software to mitigate previously identified issues in Cardiosave Hybrid and/or Cardiosave Rescue Intra-Aortic Balloon Pump devices. Consignees are asked to examine their inventory immediately to identify if they have any Cardiosave Hybrid and/or Rescue IABP devices, and to identify what software version is currently installed in devices. Devices with software revision D.00 or D.01 have been corrected. If consignee units have a software version lower than D.00 or D.01 (e.g.: B.XX or C.XX) the device still requires the software update and has not yet been corrected. If consignee units do not have the corrected software revision installed, they are to continue to follow the instructions as listed int he original recall notification until the affected device(s) has been updated. Consignees are to complete and return the provided Response Form to the provided email address. Datascope/Getinge service representatives will contact consignees to schedule installation of software, if needed. Consignees are to forward the recall notice to all device users in their facility and to any customers to whom the device was shipped.
Quantity in commerce
5,475 units
Distribution
International distribution to the countries of UNITED ARAB EMIRATES, ARGENTINA , AUSTRIA, AUSTRALIA, BELGIUM, BULGARIA, BAHRAIN, BRAZIL, CANADA, SWITZERLAND, CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, GERMANY, DENMARK, ECUADOR, EGYPT, SPAIN, FINLAND, FRANCE, UNITED KINGDOM, INDONESIA, IRELAND, INDIA, IRAQ, IRAN, ITALY, JORDAN, JAPAN, SOUTH KOREA, KUWAIT, LEBANON, LITHUANIA, LIBYA, MOROCCO, MYANMAR, MEXICO, MALAYSIA, NETHERLANDS, NORWAY, NEW ZEALAND, OMAN, PHILIPPINES, POLAND, PALESTINE, PORTUGAL, SERBIA, RUSSIAN FEDERATION, SAUDI ARABIA, SWEDEN, SINGAPORE, SLOVAKIA, THAILAND, TURKEY, TAIWAN, UNITED KINGDOM, VIETNAM, YEMEN.
Product codes and lots
Product No. 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85; UDI: 10607567112312, 10607567108407, 10607567113449; Lot No. ALL LOTS.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3135-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3134-2024Class IOngoingVoluntary: Firm initiated
Z-3135-2024Class IOngoingVoluntary: Firm initiated