Biodesign Fistula Plug C-FPS-0.4-2 - For implantation to reinforce soft tissue for repair of recto-vaginal or...
Cook Biotech, Inc. / initiated 2024-08-07 / Open, Classified / root cause: labeling and UDI
Cook Biotech, Inc. began a recall of Biodesign Fistula Plug C-FPS-0.4-2 - For implantation to reinforce soft tissue for repair of recto-vaginal or anorectal fistulas. Order Number (GPN): G46372 on . FDA classified it as Class II and gives the reason as "Products expire prior to the expiration date printed on the product labeling". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-3256-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Cook Biotech, Inc.
- Product
- Biodesign Fistula Plug C-FPS-0.4-2 - For implantation to reinforce soft tissue for repair of recto-vaginal or anorectal fistulas. Order Number (GPN): G46372
- Product code
- FTM Mesh, Surgical
- Reason for recall
- Products expire prior to the expiration date printed on the product labeling
- Root cause tag
- labeling and UDI
- FDA root cause
- Incorrect or no expiration date
- Action
- Cook Biotech Inc. issued Urgent: Medical Device Recall Letter on 8/7/24. Letter states reason for recall, health risk and action to take: 1. Examine inventory immediately to determine if you have affected product(s), and quarantine affected product(s). Immediately cease all distribution and use of Biodesign Fistula Plugs with Lot Numbers provided in the Product Information section of this letter. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Note: Unaffected products that are returned will not be credited. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email ([email protected]). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: [email protected]. Cook Biotech initiated CAPA 2024-CR-002 for this issue. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.
- Quantity in commerce
- 39 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Canada, China, Finland, France, Germany, Ireland, Luxembourg, Netherlands, New Zealand, Norway, South Africa, United Kingdom.
- Product codes and lots
- UDI: (01)00827002463721(17)250313(10)LB1524444 (01)00827002463721(17)250313(10)LB1524447 (01)00827002463721(17)250319(10)LB1524448 (01)00827002463721(17)250319(10)LB1524473 (01)00827002463721(17)250416(10)LB1525814 (01)00827002463721(17)250419(10)LB1530195 (01)00827002463721(17)250523(10)LB1534372 (01)00827002463721(17)250702(10)LB1539163 Lot Number: LB1524444 LB1524447 LB1524448 LB1524473 LB1525814 LB1530195 LB1534372 LB1539163
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3250-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3251-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3252-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3253-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3254-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3255-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3256-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3257-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3258-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3259-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3260-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3261-2024 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.