Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-3152-2024Class II

2090 Programmers with serial number prefixes PKK0 and PKK1: a) 2090, Product Description: PRGM 2090...

Medtronic, Inc. / initiated 2024-07-11 / Open, Classified / root cause: design

Medtronic Inc. began a recall of 2090 Programmers with serial number prefixes PKK0 and PKK1: a) 2090, Product Description: PRGM 2090 PROGRAMMER-PORTABLE CRM; b) 209031, Product Description... on . FDA classified it as Class II and gives the reason as "Medtronic CareLink 2090 programmers with serial number prefixes PKK0 and PKK1 have the potential for autonomous cursor motion when Finger Touchscreen capability is enabled by software version 3.2 or higher". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-3152-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
2090 Programmers with serial number prefixes PKK0 and PKK1: a) 2090, Product Description: PRGM 2090 PROGRAMMER-PORTABLE CRM; b) 209031, Product Description: PROGRAMMER 209031 PORTABLE CRM; c) 2090AA, Product Description: PROGRAMMER 2090AA PORTABLE CRM; d) 2090AA4, Product Description: 2090AA4 PROGRAMMER-PORTABLE CRM; e) 2090AC, Product Description: SFTWR 2090AC PROGRAMMER-PORTABLE CRM W/; f) 2090AM, Product Description: PROGRAMMER CARELINK HONG KONG ASIA; g) 2090CHEN, Product Description: PRGM 2090CHEN PKG-ASSY PROGRAMMER 2090; h) 2090CHFR, Product Description: PRGM 2090CHFR PKG-ASSY PROGRAMMER 2090; i) 2090CL, Product Description: PROGRAMMER 2090CL INTL-JAPAN MKT; j) 2090EN, Product Description: PRGM 2090EN PKG-ASSY PROGRAMMER 2090 UK; k) 2090FR, Product Description: PRGM 2090FR PKG-ASSY PROGRAMMER 2090 IN; l) 2090LDEN, Product Description: PRGM 2090LDEN PROGRAMMER-PORTABLE LOANE; m) 2090NL, Product Description: PRGM 2090NL PKG-ASSY PROGRAMMER 2090 IN; n) 2090W, Product Description: PROGRAMMER 2090W TELEMETRY C; o) 2090WCHDE, Product Description: PROGRAMMER 2090WCHDE SWISS-GERMAN; p) 2090WCHEN, Product Description: PROGRAMMER 2090WCHEN SWISS-ENGLISH; q) 2090WCHFR, Product Description: PROGRAMMER 2090WCHFR SWISS-FRENCH; r) 2090WDE, Product Description: PROGRAMMER 2090WDE GERMAN MKT; s) 2090WFR, Product Description: PROGRAMMER 2090WFR FRENCH MKT; t) 2090WLD, Product Description: PROG 2090WLD ALL LANGUAGES W/ANALYZER; u) 2090WNL, Product Description: PROGRAMMER 2090WNL NETHERLANDS MKT
Product code
NVZ Pulse Generator, Permanent, Implantable
Reason for recall
Medtronic CareLink 2090 programmers with serial number prefixes PKK0 and PKK1 have the potential for autonomous cursor motion when Finger Touchscreen capability is enabled by software version 3.2 or higher.
Root cause tag
design
FDA root cause
Device Design
Action
Medtronic issued a MEDICAL DEVICE CORRECTION notice to its consignees on 07/11/2024 via email or UPS 2-day delivery. The notice explained the issue, potential risk, and provided the following: "Instrument Management Recommendations Software updates are necessary to maintain proper programmer function. Medtronic representatives will assist in performing the software update on all Medtronic CareLink 2090 programmers and assess proper function. Medtronic representatives will assist with returning programmers needing repair or replacement. " Following review of this letter, please complete the enclosed Confirmation Form and email it to [email protected] to confirm receipt of this notification. " Please forward this notice to others who need to be aware within your organization."
Quantity in commerce
18788 units
Distribution
Worldwide distribution.
Product codes and lots
2090 Programmers with serial number prefixes PKK0 and PKK1: a) GTIN: 00613994136480, 00613994136497, 00613994136503, 00613994136510, 00613994136527, 00613994136534, 00613994136565, 00613994136596, 00613994136619, 00613994136626, 00613994175076, 00613994175083, 00613994175090, 00613994175106, 00613994175120, 00613994175137, 00613994175144, 00613994175151, 00613994175168, 00613994175175, 00613994175182, 00613994175205, 00613994175212, 00613994175229, 00613994175236, 00613994175243, 00613994175250, 00613994175267, 00613994175281, 00613994175298, 00613994175304, 00613994175311, 00613994175328, 00613994175335, 00613994175342, 00613994175366, 00613994175373, 00613994175380, 00613994175397, 00613994175403, 00613994175410, 00613994175434, 00613994175465, 00613994175489, 00613994175540, 00613994175588, 00613994175595, 00613994203182, 00613994360922, 00613994434883, 00613994434890, 00613994434944, 00613994435002, 00613994435255, 00613994435316, 00613994580863, 00613994580894, 00613994580931, 00613994580962, 00613994581013, 00613994581051, 00613994581075, 00613994581105, 00613994581150, 00613994581167, 00613994581174, 00613994581181, 00613994581204, 00613994581211, 00613994581228, 00613994581235, 00613994581242, 00613994581549, 00613994581570, 00613994581587, 00613994581594, 00613994581624, 00643169440128, 00643169440166, 00643169440173, 00643169440197, 00643169440203, 00643169440937, 00643169537750, 00643169537774, 00643169537811, 00643169537828, 00643169537835, 00643169537842, 00643169537859, 00643169537866, 00643169537880, 00643169537910, 00643169537927, 00643169538016, 00643169629004, 00643169629011, 00643169629028, 00643169629035, 00643169629042, 00643169629066, 00721902246227, 00721902246234, 00721902246241, 00721902246265, 00721902246289, 00721902246302, 00721902246326, 00721902246333; b) GTIN: 00721902171222; c) GTIN: 00721902171079; d) GTIN: 00721902824692; e) GTIN: 00721902059100; f) GTIN: 00721902171154; g) GTIN: 00721902246357; h) GTIN: 00721902246364; i) GTIN: 00721902480201; j) GTIN: 00721902246159, 00721902246173; k) GTIN: 00721902246203; l) GTIN: 00721902841903; m) GTIN: 00721902246272; n) GTIN: 00613994127631, 00613994435354, 00613994581549, 00885074301440, 00885074310596, 00885074310633, 00885074310657, 00885074310930, 00885074312866, 00885074312873, 00885074313047, 00885074313108, 00885074313115, 00885074313139, 00885074313146, 00885074313153, 00885074313160; o) GTIN: 00885074313191; p) GTIN: 00885074313177; q) GTIN: 00885074313184; r) GTIN: 00885074314921; s) GTIN: 00885074313009; t) GTIN: 00885074357416; u) GTIN: 00885074313122

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3152-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3152-2024Class IIOngoingVoluntary: Firm initiated