Z-3300-2024Class II
Custodiol HTK Solution, 500ml bottle. 10 per carton
Dr. Franz Kohler Chemie GmbH / initiated 2024-01-09 / Open, Classified / root cause: design
Dr Franz Koehler Chemie Gmbh began a recall of Custodiol HTK Solution, 500ml bottle. 10 per carton on . FDA classified it as Class II and gives the reason as "Potential for particles to be present in solution". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-3300-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Dr. Franz Kohler Chemie GmbH
- Product
- Custodiol HTK Solution, 500ml bottle. 10 per carton.
- Product code
- KDL Set, Perfusion, Kidney, Disposable
- Reason for recall
- Potential for particles to be present in solution.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The single distributor was notified on 01/09/2024 via email. The customer was instructed to notify customers and provide the recall letter. Any affected product should be withdrawn, and sales stopped.
- Quantity in commerce
- 48 cartons (480 bottles)
- Distribution
- US Nationwide distribution in the state of North Carolina.
- Product codes and lots
- UDI-DI: 04033133105003, Batch Numbers: 2314212, 2320012, 2322812, 2333512 (all batches)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3300-2024 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.