vanSonnenberg Sump Locking Pigtail Regular Kit Sump Catheter System Kit, Material Number REF M001273090; for...
Boston Scientific Corporation began a recall of vanSonnenberg Sump Locking Pigtail Regular Kit Sump Catheter System Kit, Material Number REF M001273090; for use as a percutaneously placed drain for... on . FDA classified it as Class II and gives the reason as "There is the potential for holes in the clear sterile barrier of the device pouch which may impact device sterility". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-3150-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Boston Scientific Corporation
- Product
- vanSonnenberg Sump Locking Pigtail Regular Kit Sump Catheter System Kit, Material Number REF M001273090; for use as a percutaneously placed drain for intra-abdominal fluid collections.
- Product code
- FFA Tube, Drainage, Suprapubic
- Reason for recall
- There is the potential for holes in the clear sterile barrier of the device pouch which may impact device sterility.
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- Boston Scientific issued am Urgent Medical Device Removal notice to it consignees on 07/18/2024 via overnight mail. The notice explained the issue, potential hazard, and requested the return of the devices. Consignees were directed to post the notice in a visible location near the affected products to ensure visibility and to share the notice with all those to whom the device was transferred. Distributors were direct to notify their customers.
- Quantity in commerce
- 4 units
- Distribution
- US, Colombia, Philippines, Russia
- Product codes and lots
- GTIN/UDI 0871472913774, Lot # 33512377, exp. 02/26/2027
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3147-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3148-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3149-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3150-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3151-2024 | Class II | Ongoing | Voluntary: Firm initiated |