Z-0009-2025Class II

BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSDuet Sample Preparation System...

Becton, Dickinson and Company, BD Biosciences / initiated 2024-08-12 / Open, Classified / root cause: software

Becton, Dickinson and Company, BD Bio Sciences began a recall of BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSDuet Sample Preparation System (Base), REF: 663128, 662588, used with software V... on . FDA classified it as Class II and gives the reason as "Sample preparation system with software may cause: 1) Multidispense feature enabled with 2-tube assay: Specimen volume dispensed into tube1 at least 3% higher than volume in tube2, or 2) Absolute counts calculated using...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0009-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Becton, Dickinson and Company, BD Biosciences
Product
BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSDuet Sample Preparation System (Base), REF: 663128, 662588, used with software V 1.4.1. Intended use: to prepare human specimens to be analyzed.
Product code
PER Automated Pipetting, Diluting And Specimen Processing Workstations For Flow Cytometric Analysis
Reason for recall
Sample preparation system with software may cause: 1) Multidispense feature enabled with 2-tube assay: Specimen volume dispensed into tube1 at least 3% higher than volume in tube2, or 2) Absolute counts calculated using Trucount Tubes: increased volume in tube1 may result in 20% higher absolute counts in tube 1 than tube 2; could result in additional blood collection, delayed results/patient therapy.
Root cause tag
software
FDA root cause
Software change control
Action
On 8/12/2024, correction notices were sent to customers who were informed of the following; 1) Cease use of the sample preparation system with software version 1.4.1. until the Multidispense feature has been disabled. Patient samples can be prepared using single dispense on the sample preparation system or manually. 2) It is recommended that laboratories upgrade affected sample preparation systems to version 1.4.2 to correct the issue as soon as it is available. 3) Ensure the contents of this notification are read and understood by those within your organization. 4) Complete and return the response form via email to [email protected] If you require further assistance, contact: 1-844-823-5433, Mon-Fri 8:00am and 5:00pm CT or email [email protected]. Recall related questions can be emailed to [email protected]
Quantity in commerce
130
Distribution
US: NJ, UT, WI, TX, ID, AK, MN, MA, MD, TN, MI, CA, VA OUS: Ireland, Austria, Italy, Germany, Belgium, Finland, Canada, France, Spain, New Zealand, United Kingdom, Japan, Australia, Norway, Singapore, Sweden, Turkey, Brazil, Colombia, Netherlands, United Arab Emirates, Saudi Arabia, Mexico, Switzerland, Costa Rica, China, Azerbaijan, Algeria, Jordan, India
Product codes and lots
REF/UDI-DI/Serial Numbers: 666340/382906663405/9230010101, 9230010102, 9230010103, 9230010104, 9230010105, 9230010107, 9230010108, 9230010109, 9230010112, 9230010113, 9230010114, 9230010115, 9230010116, 9230010118, 9230010119, 9230010120. 666339/9230000102, 9230000103, 9230000104, 9230000105, 9230000106, 9230000112, 9230000113, 9230000114, 9230000118, 9230000120, 9230000121, 9230000123, 9230000124, 9230000126, 9230000128, 9230000130, 9230000132, 9230000134, 9230000135, 9230000138, 9230000144, 9230000145, 9230000131, 9230000152, 9230000149, 9230000030, 9230000101, 9230000108, 9230000109, 9230000110, 9230000111, 9230000115, 9230000116, 9230000117, 9230000119, 9230000122, 9230000125, 9230000127, 9230000129, 9230000136, 9230000139, 9230000141, 9230000142, 9230000143. 663128/382906631282/9220030004, 9220030005, 9220030008, 9220030016, 9220030017, 9220030027, 9220030029, 9220030031, 9220030043, 9220030053, 9220030057. 662588/9220010119, 9220010179, 9220010182, 9220010232, 9220010236, 9220010250, 9220010265, 9220010275, 9220010293, 9220010311, 9220010313, 9220010316, 9220010324, 9220010327, 9220010335, 9220010338, 9220010341, 9220010344, 9220010348, 9220010349, 9220010292, 9220010112, 9220010120, 9220010125, 9220010127, 9220010128, 9220010130, 9220010133, 9220010138, 9220010154, 9220010161, 9220010183, 9220010188, 9220010197, 9220010215, 9220010225, 9220010231, 9220010238, 9220010240, 9220010242, 9220010244, 9220010256, 9220010257, 9220010260, 9220010263, 9220010268, 9220010269, 9220010279, 9220010280, 9220010289, 9220010290, 9220010294, 9220010296, 9220010329, 9220010332, 9220010345, 9220010354, 9230000031, 99220010292.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0009-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0009-2025Class IIOngoingVoluntary: Firm initiated