ADVIA Chemistry Microalbumin_2 (ALB_2) -For in vitro diagnostic use in the quantitative measurement of...
Siemens Healthcare Diagnostics, Inc. began a recall of ADVIA Chemistry Microalbumin_2 (ALB_2) -For in vitro diagnostic use in the quantitative measurement of microalbumin in human urine on ADVIA Chemistry systems/... on . FDA classified it as Class II and gives the reason as "Atellica CH Microalbumin_2 (ALB_2) and the ADVIA Chemistry Microalbumin_2 (ALB_2) are not meeting the High-Dose Hook Effect claim/Prozone Effect claim as stated in the Instructions for Use (IFU) on the Atellica CH...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-3170-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- ADVIA Chemistry Microalbumin_2 (ALB_2) -For in vitro diagnostic use in the quantitative measurement of microalbumin in human urine on ADVIA Chemistry systems/ Atellica CH and Atellica CI Analyzers.
- Product code
- JIR Indicator Method, Protein Or Albumin (Urinary, Non-Quant.)
- Reason for recall
- Atellica CH Microalbumin_2 (ALB_2) and the ADVIA Chemistry Microalbumin_2 (ALB_2) are not meeting the High-Dose Hook Effect claim/Prozone Effect claim as stated in the Instructions for Use (IFU) on the Atellica CH, Atellica CI Analyzers, ADVIA 1800 Chemistry Systems, ADVIA 2400 Chemistry Systems, and ADVIA Chemistry XPT Systems, may result in erroneously depressed microalbumin patient results.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Siemens Healthcare issued Urgent Medical Device Corrections (UMDC) to US consignees ACHC24-05.A.US, ACHC24-05.A.US.CHC on 8/15/24 via FedX and Urgent Field Safety Notices (UFSN)to foreign consignees ACHC24-05.A.OUS, ACHC24-05.A.OUS.CHC to affected customers informing them of the issue and the impact to results. ADVIA Chemistry customers are advised that lot 665399 meets the IFU prozone claim. Customers will be instructed to discontinue use of and discard the impacted reagent lots indicated in the letter. Atellica CH customers will be advised that lots 242194, 242195, and 242321 meet the IFU prozone claim. Customers can continue to use the impacted products, with the understanding that patient samples with values above 9,500 mg/dL (95,000 mg/L) can result in falsely depressed results when using the impacted lots. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Resolution Lot 665399 meets the IFU prozone effect claim. The manufacturing control system has been updated to ensure that there is no impact to future lots. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
- Quantity in commerce
- 19645 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bosnia Herzeg, Brazil, Canada, Chile, Colombia, Croatia, Cura¿ao, St. Eus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hong¿Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lithuania, Malaysia, Maldives, Mexico, Morocco, Netherlands, New¿Zealand, Norway, Oman, P.R.¿China, Pakistan, Paraguay, Peru, Poland, Portugal, Qatar, Republic¿Korea, Romania, Russian¿Fed., Saudi¿Arabia, Serbia, Singapore, Slovakia, Slovenia, South¿Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, U.A.E, United¿Kingdom, Uruguay, Vietnam, Zaire. ¿¿¿¿¿¿
- Product codes and lots
- UDI-DI: 00630414517643 Lot Numbers: 612528, 616644, 620971, 623768, 636498, 640513, 658356, 668872
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3169-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3170-2024 | Class II | Ongoing | Voluntary: Firm initiated |