Z-3309-2024Class II
Stryker Tornier Perform Humeral System, Reversed Insert, 45%, REF DWP2393, for shoulder arthroplasty
TORNIER Inc. / initiated 2024-08-29 / Open, Classified
Tornier, Inc began a recall of Stryker Tornier Perform Humeral System, Reversed Insert, 45%, REF DWP2393, for shoulder arthroplasty on . FDA classified it as Class II and gives the reason as "Devices from one lot of Perform" Reversed Inserts may contain a locking ring that was assembled upside down which would prevent the humeral insert from securely seating in the stem or spacer". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-3309-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- TORNIER Inc.
- Product
- Stryker Tornier Perform Humeral System, Reversed Insert, 45%, REF DWP2393, for shoulder arthroplasty
- Product code
- PAO Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer + Additive, Cemented
- Reason for recall
- Devices from one lot of Perform" Reversed Inserts may contain a locking ring that was assembled upside down which would prevent the humeral insert from securely seating in the stem or spacer.
- FDA root cause
- Under Investigation by firm
- Action
- Stryker issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 08/29/2024 via email. The notice explained the issue, potential risk, and requested the following actions be taken: Identify and isolate the affected units, and respond to the notice via email. Upon receipt of the response, Stryker will arrange for the return of the device. Those consignees who further distributed the product (to other sites or customers) were directed to notify those parties. For questions or concerns, contact: [email protected].
- Quantity in commerce
- 38 units
- Distribution
- US Nationwide distribution.
- Product codes and lots
- GTIN 00846832084494, Batch Lot # 1495AY
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3309-2024 | Class II | Ongoing | Voluntary: Firm initiated |