Z-3302-2024Class II

(a) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 12 Fr/Ch (4.0 mm) x 48 in. (122...

Medtronic, Inc. / initiated 2024-07-24 / Open, Classified / root cause: labeling and UDI

Cardinal Health 200, LLC began a recall of (a) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 12 Fr/Ch (4.0 mm) x 48 in. (122 cm), Model #8888266114, Sterile; (b)... on . FDA classified it as Class II and gives the reason as "The firm received reports of breakage in the Anti-Reflux Valve (ARV) due to improper use. As a result, changes to labeling were made". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-3302-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
(a) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 12 Fr/Ch (4.0 mm) x 48 in. (122 cm), Model #8888266114, Sterile; (b) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 14 Fr/Ch (4.7 mm) x 48 in. (122 cm), Model #8888266122, Sterile; (c) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 16 Fr/Ch (5.3 mm) x 48 in. (122 cm), Model #8888266130, Sterile; and (d) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 18 Fr/Ch (6.0 mm) x 48 in. (122 cm), Model #8888266148, Sterile.
Product code
FEG Tube, Double Lumen For Intestinal Decompression And/Or Intubation
Reason for recall
The firm received reports of breakage in the Anti-Reflux Valve (ARV) due to improper use. As a result, changes to labeling were made.
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
The recalling firm issued letters dated 7/22/2024 via overnight mail on 7/24/2024. The letter informed the consignee the product labeling has been updated via electronic instructions for use for the affected product due to reported incidences of improper use of the device which can lead to breakage of the Salem Sump Anti-Reflux Valve (ARV) and an increased risk to patients. To mitigate the risk of improper use, the instructions have been revised. The risk to health and actions required were included in the letter. The actions included: (1) Review your inventory for the affected product codes; (2) Communicate the change of the use instructions with all personnel that utilize the devices; (3) Post a copy of Attachment A (the revised IFU) and the recall notification in the storeroom and clinical areas; (4) Notify any customers to whom you may have distributed/forwarded affected product or to whom you intend to distribute/forward product) regarding this medical device product correction and share a copy of the notice; (5) Return the enclosed acknowledgment form via FAX or email.
Quantity in commerce
24,761,100 tubes total
Distribution
Distribution was nationwide, including Puerto Rico and Guam. There was also government/military distribution. Foreign distribution was made to Chile.
Product codes and lots
All lot numbers. UDI-DI numbers: (a) 8888266114, UDI-DI 10192253012781; (b) 8888266122, UDI-DI 10192253012804; (c) 8888266130, UDI-DI 10192253012828; and (d) 8888266148, UDI-DI 10192253012842.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3302-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3301-2024Class IIOngoingVoluntary: Firm initiated
Z-3302-2024Class IIOngoingVoluntary: Firm initiated