CADD Medication Cassette Reservoir, 100mL, FS, YELLOW, Product Code 21-7300-24
Smiths Medical ASD Inc. began a recall of CADD Medication Cassette Reservoir, 100mL, FS, YELLOW, Product Code 21-7300-24 on . FDA classified it as Class II and gives the reason as "Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0012-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Abbott Medical
- Product
- CADD Medication Cassette Reservoir, 100mL, FS, YELLOW, Product Code 21-7300-24
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Smiths Medical issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 06/26/2024 via mail. The notice explained the issue, potential risks, and requested the affected product be quarantined, and the response for returned to initiate the return and replacement process. Distributors were directed to notify their customers. For further inquires: Product complaints or adverse events: [email protected] 1-(866)-216-8806 Additional information or technical assistance 1-(800)-241-4002, option 3 [email protected]
- Quantity in commerce
- 26160 units
- Distribution
- Worldwide distribution.
- Product codes and lots
- UDI/DI 10610586027192, Lot Numbers: 3971732, 3971731, 4034016, 3977457, 4013365, 4019421, 4019422, 4034015, 4037744, 4037745, 4034018, 4034017, 4046844, 4046843, 4053912, 4048899
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0010-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0011-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0012-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0013-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0014-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0015-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0016-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0017-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0018-2025 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.